156
156
0 - 40 mg/l
(MoO )
4
Name on device: 134-Molybd-HR-pp
(134)
Molybdate HR
Powder
Continued...
Note(s)
-> The following substances interfere with the measurement: Aluminium
(> 50 mg/l), Chromium (> 1000 mg/l), Iron (> 50 mg/l), Nickel (> 50 mg/l),
Nitrite (>2000 as NO2; can be eliminated by adding one Sulfamic Acid
powder pillow to the sample).
-> In case sample contains >10 mg/l Copper, increases positive reading if test
is not performed quickly enough before pressing TEST.
-> Highly buffered samples or samples with extreme pH levels may require pre-
treatment.