18) Proceed to next step to intitiate next countdown.
19) Reagents need to react press TEST to start 02:00 minutes countdown.
20) Open 1 Acid LR/HR vial(s) and add 2ml from ZERO vial to one.
This is your new ZERO vial.
21) Add 2ml from former TEST vial to second new vial. This is your new
SAMPLE vial.
22) Cap both cells and shake to mix. CAUTION: heat build-up!
23) Place the 16 mm adaptor into the device.
25) Place the ZERO cell and the light protection cover and than press
ZERO.
26) Remove the cell from the device.
27) Place the sample cell and the light protection cover into the device.
Than press TEST.
B) Only ZERO and TEST:
1) Place the 16mm adaptor into the device.
2) Place the ZERO cell and the light protection cover and than press
ZERO.
3) Remove the cell from the device.
4) Place the sample cell and the light protection cover into the device.
Than press TEST.
(151)
Nitrogen-Total (LR)
0.5 - 25 mg/l
(N)
Name on device: 151-NitroTotLR-pre
Continued...
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 22)
-> Use last ZERO
= continue with step 25)
1
2
3
4
16mm vial
4
4
4
4
4
Continued...
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...