179
179
Tablet
Continued...
Note(s)
-> If the sample water contains further reducing chemicals (e.g. chlorine,
activated oxygen, bromine etc.) this will also be detected and is included in
the test results.
-> When preparing the measurement procedure it is important to avoid any
chlorine escaping, which can happen during pipetting and shaking the
sample. The measurement should be performed directly after sampling.
-> It is important that the measurement devices to be used have not been
cleaned with household detergent, as this would greatly influence the
measurement. To prevent any contamination the cuvette, the cuvette lid and
the stirrer should be stored in a 0.1 % sodium hypochlorite solution for one
hour and then thoroughly rinsed with distilled water.
-> The DPD reagent used in this procedure buffers the pH value of the sample
water in the range between 6.2 and 6.5 pH. If your sample water is very
alkaline or acidic this must be adjusted to a pH range between 6 and 7 by
the addition of 0.5 mol/l sulphuric acid or resp. 1 mol/l caustic soda before
the DPD reagent is added.
Name on device: 37-Ozone-tab
(37)
Ozone
0 - 5.4 mg/l
(O )
3
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...