211
211
0 - 20 mg/l
Name on device: 87-Phosphonate-liq
1) Fill 8 ml distilled water into a clean 24mm cuvette.
2) Add precisely 2 ml sample water to the same cuvette.
3) Replace the lid on the cuvette and swivel this back and forth 5 times.
4) Replace the cuvette in the sampling chamber in the PrimeLab. Make
sure the arrow on the front of the cuvette is aligned. Press the ZERO
button.
5) Remove the cuvette from the chamber and unscrew the lid.
6) Add 1 x 0.05ml (scoop) "PL Phosphonate 1" powder to the sample
water .
7) Replace the lid. Swivel the cuvette to mix the liquid and the reagent
until this is competely dissolved. Press the TEST button and wait for the
end of the 5 minute countdown. The cuvette does not have to be inside
the sampling chamber at this time.
8) Remove the cuvette from the chamber and unscrew the lid.
9) Add 1 x 0.05ml (scoop) "PL Phosphonate 2" powder to the sample
water .
10) Replace the lid. Swivel the cuvette to mix the liquid and the reagent
until this is competely dissolved. Press the TEST button and wait for the
end of the 2 minute countdown. The cuvette does not have to be inside
the sampling chamber at this time.
(87)
Phosphonate
4
4
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 2)
-> Use last ZERO
= continue with step 5)
1
2
3
4
Liquid
+Powder
10 ml
24 mm
PL Phosphonate 1
(PLpow20PPHON1)
PL Phosphonate 2
(PLpow20PPHON2)
PL Phosphonate 3
(PL65PPHON3)
PL Phosphonate 4
(PLpow20PPHON4)
Powder
4
Filter
GF/C
(*)
(*)
(*)
Syringe
10
9
8
7
6
5
4
3
2
1
ml
Filter paper Filter adapter
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...