PTSA
0 - 1000 µg/l
(PTSA)
Name on device: 111-PTSA-Ad
Adapter
Continued...
Notes
->
Turbidity in samples may affect the PTSA result. Filter any turbid
samples using GF/C filter paper before commencing PTSA
measurement.
->
Ensure that all parts are clean, dry and free of grease and the adapter
must be placed firmly until it stops.
-> As this Parameter uses an indirect light from above, the shape of the bottom
of the cell is important to the result. Because the bottom of the cells can vary
greatly it is imperative to always use the cell with which this parameter was
also calibrated. It is essential to always ensure the correct amount of water
in the cell, which is why exactly 10 ml of liquid should be taken by the
pipette for the subsequent sample measurement. Please change or clean
the tip of the pipette after each measurement/calibration.
-> Calibrate this parameter via the calibration SET if you use another cell or
you have the feel that the measurement result is inaccurate.
-> One of the following reasons can lead into receiving an error message:
“check adapter”
• Weak or empty batteries (please change)
• Dirty lense (adapter)
• Wrong adapter used for this measurement (there are different adapter for
different measurements, all looking the same)
• Adapter might not stay straight on PrimeLab
• Cuvette-hole (PrimeLab) might be dirty (check the two windows)
• Water sample might be too dark / not enough light can pass water sample
to reach the sensor
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...