258
258
(117)
Transmission-570nm
0 - 100 %
(Trnsm)
10 ml
Name on device: 117-Transm-570nm
24 mm
Continued...
1) Fill 10ml of the sample water into a clean 24mm cuvette.
2) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
ZERO.
3) Remove the cuvette from the chamber and unscrew the lid.
4) Treat water sample according to the method you did chose.
5) Replace the cuvette in the sampling chamber in the PrimeLab. Make
sure the arrow on the front of the cuvette is aligned. Press the TEST
button.
6) Ihnen wird nun unmittelbar das ermittelte Ergebnis angezeigt. Mit
Drücken der Taste „OK“ wird der ermittelte Wert nebst Datum und Uhrzeit
im Gerät gespeichert.
4
4
4
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 2)
-> Use last ZERO
= continue with step 4)
1
2
3
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...