Note(s)
-> (*) conversion to FTU (same as FAU)
-> FAU stands for Formazin Absorption Units, different from the NTU
(nephelometric) method.
-> The measurement should be conducted immediately after sampling.
-> Air bubbles will influence the measurement result.
-> Tinted water samples influence the result. In this case do not use distilled
water (step 1) but rather filtered sample water for the ZERO adjustment.
-> The Turbidity test measures the optical value of the sample which results
from the scattering and absorption of light particles.
The amount of turbidity depends on variables such as size, shape, colour
and the refractive nature of the particles.
This test is calibrated using Formazin Turbidity Standards and the readings
are in terms of FAU (Formazin Attenuation Units).
This test can be used for daily plant monitoring and 1 FAU is equivalent to 1
NTU (Nephelometric Turbidity Unit).
This test is not suitable for USEPA reporting purposes as the optical method
of measurement for FAU is very different than the NTU method.
However 1 NTU = 1 FTU = 1 FAU when traced to formazin primary
standards.
20 - 1000 FAU
Name on device: 59-Turbidity
(59)
Turbidity
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...