0 - 1000 µg/l
(Watch)
1) Rinse vial 3 times with your water sample.
2) Fill a cleaned cell (24mm) with 10ml of your sample.
3) Place the cell into the PrimeLab (use caution to the arrow on the cell).
Do NOT close the lid.
4) Place PTSA-ADAPTER on top of the open cuvette.
5) Press TEST.
10 ml
Name on device: 156-Watch-Ad
24 mm
4
Adapter
PLSp-PIP10ml
PLSp-PIP10ml-tips
PLSp-ADP-PTSA
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
Note(s)
->
Ensure that all parts are clean, dry and free of grease and the adapter
must be placed firmly until it stops.
1. As this Parameter uses an indirect light from above, the shape of the bottom of
the cell is important to the result. Because the bottom of the cells can vary
greatly it is imperative to always use the cell with which this parameter was also
calibrated. It is essential to always ensure the correct amount of water in the cell,
which is why exactly 10 ml of liquid should be taken by the pipette for the
subsequent sample measurement. Please Change or clean the tip of the pipette
after each measurement/calibration.
2. Calibrate this parameter via the calibration SET if you use another cell or you
have the feel that the measurement result is inaccurate.
3. One of the following reasons can lead into receiving an error message: “check
adapter”
• Weak or empty batteries (please change)
• Dirty lense (adapter)
• Wrong adapter used for this measurement (there are different adapter for
different measurements, all looking the same)
• Adapter might not stay straight on PrimeLab
• Cuvette-hole (PrimeLab) might be dirty (check the two windows)
• Water sample might be too dark / not enough light can pass water sample
to reach the sensor
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...