63
63
0 - 1500 mg/l
(O )
2
Name on device: 80-COD-MR-pre
(80)
COD MR
COD-80-MR
1) Prepare two cells (16mm) "COD-80-MR". Label one as zero cell.
2) Open the first cell (zero) and fill with 2 ml deionised water.
3) Open the second cell (sample) and fill with 2 ml of your sample water.
4) Cap both cells and shake to mix. CAUTION: Cells turn hot!
5) Put the cells in a preheated thermoreactor for 120 minutes at 150°C.
6) CAUTION: cells are hot! Remove the cells from the thermoreactor and
allow to cool to 60°C or less.
7) While warm invert cells several times to mix the contents well. Allow to
cool to room temperature.
8) Place the prepared vial adaptor into the device.
9) Place the zero cell and the light protection cover into the device and
then press ZERO.
10) Remove the cell from the device.
11) Place the sample cell and the light protection cover into the device.
Then press TEST.
4
Continued...
The individual steps of the measurement procedure are shown on the display.
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
!
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5".
Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
One-Time-Zero:
Measure new ZERO
Use last ZERO
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
!
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
= continue with step 9)
-> Use last ZERO
= continue with step 11)
1
2
3
4
4
16 mm cuvette
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...