64
64
0 - 1500 mg/l
(O )
2
(80)
COD MR
12) The determined result is immediately displayed. Press the "OK"
button to save the result together with date and time.
Continued...
Note(s)
-> Suspended particles in the zero cell and / or the sample cell lead to wrong
test results. Make sure that any existing suspended solids have settled to the
bottom of the cell and are not disturbed by the insertion into the PrimeLab.
-> Both cells used for the measurement (zero / sample) must be from the same
production batch. The cell used for ZERO can be kept for other tests (of the
same batch) but must be stored in the dark.
-> For COD content above 1500 mg/l, the use of another method (COD HR) is
recommended to achieve accurate results.
-> Never insert hot cells into the PrimeLab measuring chamber!
-> This method is not suitable for water samples with Chloride values higher
than 1,000 mg/l.
4
!
Possible messages in the line below the measurement value:
"low! / high! / good!":
Assessment of the measurement value relative to the ideal range defined by you.
"Overrange! / underrange!":
Measurement value for this parameter is outside the value range of this method.
Name on device: 80-COD-MR-pre
16 mm cuvette
Summary of Contents for 1.0 MULTITEST
Page 1: ......
Page 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Page 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Page 33: ...left open for technical reasons ...
Page 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Page 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Page 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Page 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Page 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Page 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...