8.2.1
Align the Cube L1 with the alignment tool
Using the alignment tool and a pilot laser beam, you can position the device with the
necessary accuracy (see Fig. 8.2 on page 16):
1.
Put the alignment tool on the inlet aperture.
2.
Turn on the pilot laser and align the device:
•
If the laser beam hits the little marking in the alignment tool, the device is aligned.
NOTICE
Damaging/Destruction of the device
The exposure with laser radiation will destroy the alignment tool and the
underlying protective window.
X
Remove the alignment tool before measurement.
Marking
Alignment tool
Fig. 8.2: Alignment of the Cube L1 using the alignment tool