PRiO User’s Guide
98
Disinfecting the PRiO Device and Accessories for Multiple Users
Note:
This section applies only when more than one person will be using the same unit. In that situation,
it is vital to effectively disinfect the unit and accessories that have been used by one person before
allowing them to be used by another.
First Step: Put on protective gloves
Next Step: Clean the PRiO components and accessories
1.
Wipe down the PRiO components and accessories with disinfectant wipes (Virucidal, Bactericidal,
Pseudomonacidal, Tuberculocidal, Fungicidal)
—
Metrex CaviWipes™. Follow the disinfectant
product manufacturer’s instructions for
cleaning
.
2.
Take a Q-tip with cleaner applied (for example, Windex® or another commercial cleaner) and clean
inside all cracks and crevices. You may want to use a toothbrush or similar brush.
3.
Blow the components off with an air hose or wipe it dry.
Important! Make sure all soil is removed from the PRiO components and accessories. This is vital
before proceeding to disinfecting the equipment.
Next Step: Disinfect the PRiO components and accessories
1.
Wipe down the PRiO components and accessories with a new disinfectant wipe (Virucidal,
Bactericidal, Pseudomonacidal, Tuberculocidal, Fungicidal)—Metrex CaviWipes. Follow the
disinfectant product manufacturer’s instructions for
disinfecting
.
2.
Make sure to clean inside all cracks and corners, and wipe more than once if necessary to keep the
components wet for a minimum of three (3) minutes.
3.
Allow the components and accessories to dry.
Final Step: Wipe the touchscreen
1.
When the PRiO components and accessories are dry, wipe down the viewing area (device
touchscreen, keyguard, etc.) with glass cleaner, so the screen does not discolor.
2.
Allow the equipment to dry.