280490 FGSH-450
- Owner’s Manual
Rev: 03/25/13
Page: 18
Basics in transferring an individual… continued
Caution: Prior to lifting an individual make sure that the straps of the sling are securely
placed on the sling hooks of the carry bar.
Step 4)
Attach the straps of the sling to the sling hooks of the carry bar. The straps on each side of the sling are
generally attached to the corresponding side of the carry bar. Be sure to double check to ensure that the straps
are properly attached to the carry bar, and that the individual being lifted is properly positioned in the sling pri-
or to lifting.
Chain Hooks
Sling Strap
Hooks
Sling Strap
Hooks
Caution: Attach sling straps to the sling hooks only. Do not attach sling straps to chain
hooks. Failure to comply with this caution could result in serious injury to the
individual being lifted and/or damage to the lift.
Step 5)
The individual may now be raised by pumping the lift handle of hydraulic actuator. While lifting is in
progress the height required in order for the transfer to be completed safely should be closely observed.
Caution: When elevated few inches off the bed and before moving the patient, check that
sling straps are properly attached. If any attachments are not properly in place,
lower the individual back onto the bed and correct the attachment.
Adjustments for safety and comfort should be made before moving the individual. Patient’s
arms should be inside of the sling straps.
Prism Medical Slings and hardware are made specifically for use with Prism Medical Patient
Lifts. For the safety of the patient, DO NOT intermix slings and patient lifts of different manu-
facturers.