If GEN2 is selected, the Channel Status can only be set to a simple Consumer or
Professional pattern. The display format is:
Ù
Status:Profesnl
The Professional and Consumer patterns are:
Professional:
Audio use;
No emphasis;
Source locked;
fs as CARRIER panel indication;
Stereo;
No user data;
24-bit samples;
Not a reference;
No source or destination IDs;
No timecodes;
All bytes reliable;
CRC correct;
Consumer:
Audio use;
No emphasis;
Copy permitted;
2-channel;
Mode-0;
Category code: General;
No source or destination numbers;
fs as CARRIER panel indication, grade II;
To return to the `Generator' level of the menu system, press [EXIT].
6.4.6.
C-S Capture (menu 4.6.)
On entering the C-S Capture sub-menu, the DSA-1 checks that the currently selected
DI is locked (i.e. the CARRIER panel 'FAIL' LED is not lit). If so, the following message
is displayed:
Capturing
Channel Status
This announces that the Channel Status pattern currently on the selected DI input has
been copied into the Channel Status edit buffer for GEN1, and that each Channel
Status field has been set to 'Edit' as described in the previous section). This feature