8
USING YOUR NEBULIZER
NOTE: Before the first use of the nebulizer or if the unit has not been used for an extended
period of time, it should be cleaned thoroughly as described in the
Cleaning
section of this
manual.
WARNING: Before connecting the power cord, make sure the I/O (ON/OFF) switch is in the
O (OFF) position.
The plug is also served to disconnect the device.
CAUTION: The Compressor is designed for intermittent use only. Do not operate it
continuously for more than 20 minutes for a single use without turning it off and following a
cooling period for least 40 minutes.
Note: the compressor pump motor has a thermal protector which will turn the power off
before the unit overheats. If the thermal protector shuts the power off:
1.
Switch the power switch to the OF (O) position
2.
Unplug the power cord from the electrical outlet
3.
Make certain the air slots on the unit are not obstructed
4.
Wait 40min for the motor to cool
5.
Restart the compressor
6.
If you are still experiencing problem, contact your healthcare dealer for
assistance