background image

15

 

© 2012 by Proceq SA

4.9.3 

Measuring with the Wheel Electrode

As described in 4.9.2, begin at the location corresponding to the top left hand corner of the screen 
and move in the direction indicated or alter this accordingly.

• 

Press the START button to acquire the first reading.

• 

Now proceed to move the wheel electrode in the Y-direction (down) as indicated by the direction 
arrow (Field 2 in fig 4.1). The measurement readings will be acquired automatically in the pre-
selected row spacing.

• 

When you come to the end of a column, press the 

 button to indicate this and the cursor will 

jump right to the next column. The direction arrow will automatically change direction to 

.

• 

Now place the wheel physically one column spacing to the right and acquire the first measure-
ment by pressing the START button. Proceed to measure back along the Y-axis (up) as indicated 
by the direction arrow.

• 

When you come to the end of the measurement surface, (the X-axis) the cursor will automatically 
jump to the next column and the direction arrow will change back to

 ↓

.

• 

Following each manual or automatic change in the direction arrow, the path measurement and 
the automatic acquisition of the potential reading is halted.

• 

These breaks in the measurement process can be used for example to refill the water compart-
ment. Automatic measurement begins again when the START button is pressed.

• 

During such breaks, the measured value is shown in the status row. (Field 3 in Fig 4.1)

• 

The maximum speed of travel should not exceed a maximum of 1m/sec.

• 

Measurements along the X-axis cannot be acquired. (Objects may be rotated later in ProVista)

i

Note:

 The nozzle (Ø 1 mm) is fit as standard. On dry surfaces, the felt hoop should 

leave a clearly visible trace as you measure. Depending on the character of the 
concrete it may be necessary to increase to one of the larger nozzle diameters (Ø 
2 or 3 mm).

i

Note:

 On dry, vertical surfaces, measurements should be made downwards as this 

ensures an even moistening of the surface.

4.9.4 

Re-opening an object

An object can be re-opened and missing measurements can be entered or suspect measurements 
can be overwritten. However the following points should be observed.
Once an object has been closed and a new object has been opened, it is no longer possible to add 
pages to a previously opened object.
If you have worked out the size of the area assigned to an object (see 4.1), it is possible to reserve 
the number of pages you will require in the case that you are not able to complete the measurement 
in one go.
A page is reserved once it has appeared on the screen. In order to reserve the necessary pages, 
simply move the cursor to each page you will require. For a better orientation it may help to place an 
“X” (see Note in 4.9.2) somewhere on the page (but not in the top left hand corner). 
The last object that was opened can be re-opened and extended at any time.

4.9.5 

Overwriting or deleting readings

Existing readings can be overwritten by moving the cursor to the appropriate point and making a 
new reading with the electrode. Once the reading has registered, the cursor will move to the next 
measuring point along the axis indicated by the direction arrow.
To delete an existing reading, move the cursor to the point you wish to delete and press PRINT for 
two seconds. Once the reading has been deleted, the cursor will move to the next measuring point 
along the axis indicated by the direction arrow.

Summary of Contents for canin+

Page 1: ...canin more than 50 years of know how you can measure Made in Switzerland Operating Instructions Corrosion Analyzing Instrument...

Page 2: ...robe to the indicating device 8 3 3 Power ON and select the correct operating mode 9 3 4 Functional check of the equipment 10 4 Real Tests Potential Measurements 10 4 1 Planning and preparation 11 4 2...

Page 3: ...uency 27 6 12 Chipping graph 28 6 13 Annotations 28 7 Measuring the electrical resistivity 29 7 1 Preparing the concrete surface for measurement 29 7 2 Reading the display screen 30 7 3 Measuring with...

Page 4: ...correct Proceq SA makes no warranties and excludes all liability as to the completeness and or accuracy of the information 1 3 Safety Instructions The equipment is not allowed to be operated by childr...

Page 5: ...they form hydroxide OH with the generally available water and oxygen This principle creates a potential difference that can be measured by the half cell method The basic idea of the potential field m...

Page 6: ...ments Electrical resistivity of the concrete cover Measure with the Wenner Probe This topic is dealt with in detail in section 2 4 Oxygen content at the reinforcement With decreasing oxygen concentrat...

Page 7: ...rate of cor rosion decreases 2 5 Influence of Rebars on Electrical Resisitivity Measurements The presence of rebars disturbs electrical resisitivity measurements as they conduct current much better th...

Page 8: ...lphate with 100 units by weight of distilled water In order to ensure that the solution remains saturated add an additional teaspoonful of copper sulphate crystals into the electrode The electrode sho...

Page 9: ...For potential measurements the instrument must be in Corrosion analys ing mode Switch between modes by pressing MENU place the cursor at Wenner Probe press START select OFF for Corrosion Analysing mo...

Page 10: ...ect measurement on the concrete surface possible Yes No Can the influence from surface coatings be estimated and compensated Yes No Remove the surface coating 4 8 4 9 Pre moisten the surface if requir...

Page 11: ...y be used For more accessible horizontal vertical or overhead surfaces it is much faster to use the 1 Wheel Electrode For large horizontal surfaces the 4 Wheel Electrode is recommended for the reason...

Page 12: ...often used for example on the walls and ceiling of underground car parks however this can cause a small shift of the potentials It is always necessary to check whether or not a measurement can be mad...

Page 13: ...ry The number of pages that are still available for use is shown in the top right hand corner of the measuring screen Fig 5 1 e g 491 P indicates 491 pages are still available The number of pages cont...

Page 14: ...the measurement it should be possible to see a wet patch when applied on dry con crete If this is not the case then the foam rubber plug should be moistened with water again The simplest way to procee...

Page 15: ...ista i Note The nozzle 1 mm is fit as standard On dry surfaces the felt hoop should leave a clearly visible trace as you measure Depending on the character of the concrete it may be necessary to incre...

Page 16: ...Example of a typical distribution If the surface under test has both actively corroding as well as passive rebars then the two states exhibit differing statistical distributions of the potential In t...

Page 17: ...using a small grid is recommended to deter mine the actual Hotspot local potential minimum 5 General Settings 5 1 Backlight The display has a back light which can be switched on or off by pressing the...

Page 18: ...U END MENU START MENU END OFF ON END END START Main menu for resistivity measuring Pressing MENU saves the settings and takes you to the next higher menu Pressing END saves the settings and takes you...

Page 19: ...Pressing MENU saves the settings and takes you to the next higher menu Pressing END saves the settings and takes you to the measuring screen Use the and to select navigate and adjust settings Fig 5 2...

Page 20: ...alue has stabilized If automatic measuring is not selected then values have to be acquired by pressing the START button In both cases values between 200 and 50mV have to be acquired by pressing the ST...

Page 21: ...hese graphs can be inserted in an assessment report and serve the corrosion engineer together with results of non destructive and destructive tests like concrete cover depth carbonation depth chloride...

Page 22: ...Output Fig 5 2 in the Canin display device and press START Click on VistaTransfer button to start the download programme A dialog box appears see Fig 6 2 Select the respective COM port All the object...

Page 23: ...It may be that the objects from a com plete site were not all measured in the same direction See 4 9 3 ProVista allows this to be corrected Select the correct orientation and press OK again to display...

Page 24: ...ion Click on Configuration to alter the appearance and the language of the potential map Fig 6 5 Configuration Window The number range to which the colours are assigned can be changed through varying...

Page 25: ...le button and select the file you wish to add If the file is valid the next screen Fig 6 6 will allow the file to be correctly positioned The new file may be rotated or mirrored to match its orientati...

Page 26: ...ific value 6 9 Additional Functions Save File to save your changes Print View prints the currently visible section of the potential map respectively chipping graph Print File prints the entire measure...

Page 27: ...iagram 6 11 Cumulative frequency Fig 6 9 Cumulative frequency diagram Following an evaluation See 4 10 the corrosion engineer can place the 3 cursors in such a way that they represent areas of differe...

Page 28: ...ture like measured chloride content cracks or other defects can be placed directly in the potential map or the chipping graph The information will be printed and shown in bitmaps By default they are c...

Page 29: ...mmended that annotations within one file be added at different dimension levels i e zoom levels 7 Measuring the electrical resistivity Fig 7 1 Wenner four point probe with cable and control plate The...

Page 30: ...same location moving the probe a few mm between each measurement and taking a median from the 5 values The well moistened probe should be pressed lightly against the surface of the concrete until the...

Page 31: ...f the indicating device In case your PC hardware doesn t support serial ports you can use the USB converter part no 390 00 542 Preparing the PC for the Data Transfer Create directory C PROCEQ Windows...

Page 32: ...surements 980 pages 240 measurements each organised in up to 71 objects and 5 800 resistivity measurements 24 object files tables 256 measurements each Data Output RS 232 interface with USB adapter Ba...

Page 33: ...trode cable 1 5m 4 9 ft cable coil 25m 82 ft CANIN ProVista PC software on memory stick bottle with copper sul phate 250g 330 00 205 Canin Configuration with Rod and Wheel Electrodes Basic equipment s...

Page 34: ...5 Cable clip for telescopic extension 380 02 508 S Foam pad to resistance probe set of 4 pcs 11 Maintenance and Support 11 1 Functional check of the electrodes Correct function can be checked by use o...

Page 35: ...wooden plug facing upwards 11 4 Functional check of the resistivity probe See section 5 5 Assure that the Device Constant shown on the Indicating Device corresponds to the code engraved on the probe...

Page 36: ...asia proceq com Proceq Rus LLC Ul Optikov 4 korp 2 lit A Office 412 197374 St Petersburg Russia Phone Fax 7 812 448 35 00 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharjah Unite...

Reviews: