32
© 2012 by Proceq SA
Start of Transfer
•
Switch on the Indicating Device.
•
Press “MENU” and select “Data Output”
•
Select “Data to PC” as described in the operation instructions of the instrument.
•
Start by pressing the button “START”.
Data Storage in the PC
•
Choose in the menu “Transfer/capture text/stop”.
The data are stored in the file Data.txt and are ready for further processing.
Saving the HyperTerminal Settings
•
Choose “File/Exit” in the menu.
•
Answer the question “You are currently connected, you want to disconnect now?” with “Yes”.
•
Answer the question “Do you want to save the connection named PROCEQ ?” with “Yes”.
Display and Processing the Data
The text file “Data.txt” in the folder “C:\PROCEQ” can be displayed with any text editor or word
processing program. For editing and further processing of the data we recommend MS Excel.
9 Technical Specifications
General
Temperature range:
0° to 60°C
Display:
128 x 128 pixel graphic LCD with backlight
Impedance:
10 M
Ω
Memory:
Non-volatile memory for simultaneous storage of up to 235’000
potential measurements (980 pages @ 240 measurements each
organised in up to 71 objects) and 5’800 resistivity measurements (24
object files / tables @ 256 measurements each)
Data Output:
RS 232 interface, with USB adapter
Battery Operation:
Six LR 6 batteries, 1.5V for up to:
- 60 hours (or 30 hours with activated backlight) during potential
measurement
- 40 hours (or 20 hours with activated backlight) during resistivity
measurement
Case Dimensions:
580x480x210 mm (22.8”x18.9”x8.3”)
Weight:
Net. 10.6 kg (23.5 lbs); Shipping 14kg (31.1 lbs)
(with rod and 1-Wheel Electrodes as well as Wenner Probe)
Potential Measurement
Measurement range:
-999 mV to +340 mV
Resolution:
1mV
Electrodes:
Rod Electrode (copper/copper sulphate)
1- and 4-Wheel electrode systems (copper/copper sulphate) with
telescopic handle, integrated path measurement and water reservoir
Data Transfer:
CANIN ProVista software for downloading data and evaluation on
PC
Resistivity Measurement
Measurement range:
0 to 99 k
Ω
cm
Resolution:
1 k
Ω
cm
Resistivity probe:
Resistivity probe with integrated electronics for measuring the spe-
cific resistivity using the four-point method.
Nominal current:
180μA
Frequency:
72 Hz
Data Transfer:
by Windows Hyperterminal for analysis with third party software (e.g.