Page 30 | 49
8.5
Parameter table for the local display
Table 4: Parameter table
8.5.1
Parameter 1: Display scale
The instrument will show the measurement signal
D3 = 0 in 000...100 [%] of Range with no decimal point (default)
D3 = 1 in 04.00 … 20.00 [mA] with a fixed decimal point, 2 digits before and after the decimal point.
If the 4-20mA loop is electrically not properly connected and therefore not closed,
the display will show
OL
(open Loop) to indicate this status.
D3 = 2 in 99.99 … 9999 [mg/m3] with a floating decimal point 99.99[mg/m3] or 999.9[mg/m
3
] or
9999[mg/m
3
]
To change the display scale, follow the below procedure
Navigate to Parameter 1
Key the desired value (1, 2 or 3) into the instrument.
Safe the result by pressing K3
From now on the instrument will show the measurement result in the desired unit.
8.5.2
Parameter 2: Averaging time [sec]:
The averaging (running average) time has been adjusted to 50 seconds in the factory but can be adjusted
freely between 0 and 6000 seconds (1 h 30 min).
No.
Description
Display
303 304
1
Display scale
D3 = 0 = 0.0...100.0 [%] of Range, default
x
x
D3 = 1 = [mA]
x
x
D3 = 2 = [mg/m
3
], shows - - - - if mg/m
3
is disabled
x
2
Averaging time [sec]
000...300 [sec], default = 50 [sec]
x
x
3
20 mA scaling (Range setting)
000…999 [%], default 100%
determined by the value after Auto setup (20 times Baseline)
x
x
4
Alarm limit A [%] of Range
000...100 [%] of Range, default = 25% of Range (8mA)
x
x
5
Alarm limit B [%] of Range
000...100 [%] of Range, default = 100% of Range (20mA)
x
x
6
Alarm delay time [sec]
000...180 [sec], default = 30 [sec]
x
x
7
Flow velocity in [m/s] at 4 [mA]
000...99.9 [m/s], default = 0 [m/s]
x
8
Flow velocity in [m/s] at 20 [mA]
000...99.9 [m/s], default = 0 [m/s], no compensation
x
9
Zero & Span check interval
000...999 [hour], default = 0 [hour], no check
x
0
Command parameter
001 Reset to factory defaults
x
x
002 Enable mg/m
3
calibration
x
003 Disable mg/m
3
calibration
x
1.
Display Intercept
a
(Integer) to show [mg/m
3
] -99...099 [mg/m
3
], default = 0 [mg/m
3
]
x
2.
Display Intercept
a
(Decimal) to show [mg/m
3
] 000...999 [mg/m
3
], default = 0 [mg/m
3
]
x
3.
Display Slope
b
(Integer) to show [mg/m
3
]
000...999 [mg/m
3
/mA], default = 0 [mg/m
3
/mA]
x
4.
Display Slope
b
(Decimal) to show [mg/m
3
]
000...999 [mg/m
3
/mA], default = 0 [mg/m
3
/mA]
x
5.
20 mA scaling in 000…999 [IEU] x 1 000 000
20 mA scaling in 000…999 [IEU] x 1 000 000
x
6.
20 mA scaling in 000…999 [IEU] x 1 000
20 mA scaling in 000…999 [IEU] x 1 000
x
7.
Info: Firmware Version
Info: Firmware Version
x
x
Summary of Contents for Sintrol Dumo
Page 45: ...Page 45 49 15 Appendix 15 1 ISO 9001 certificate ...
Page 46: ...Page 46 49 15 2 MODBUS RTU register map ...
Page 47: ...Page 47 49 ...
Page 48: ...Page 48 49 ...