PROFIBUS USER MANUAL Leine & Linde AB
Part Id: 740590-01
36
Document Id: 740590 Ver. 01
Publication date: 2010-06-01
DDLM_Slave_Diag
Octet
28
29
30
31
Bits
31 – 24
23 – 16
15 – 8
7 – 0
Data
231 – 224
223 – 216
215 – 28
27 – 20
Operating
time
Table 33 Diagnostics, operating time
5.6.14 Offset Value
The offset value is calculated by the preset function and shifts the position value by
the calculated value. The offset value is stored in the device and can be provided by
the diagnostic octets 32 to 35. The data type for the offset value is a 32-bit binary
value with sign, whereby the offset value range is equal to the measuring range of the
device.
The preset function is used after the scaling function. This means that the offset
value is indicated according to the scaled resolution of the device.
DDLM_Slave_Diag
Octet
32
33
34
35
Bits
31 – 24
23 – 16
15 – 8
7 – 0
Data
231 – 224
223 – 216
215– 28
27 – 20
Offset
Value
Table 34 Diagnostics, offset value
5.6.15 Offset Value of the Encoder Manufacturer
The manufacturer offset value indicates the encoder offset set by the manufacturer.
This value gives information on the shift of the position zero point in number of
positions from the physical zero point of the encoder. The data type for the offset
value is a 32-bit binary value with sign. The value range corresponds to the measuring
range of the encoder. The offset value of the manufacturer of the encoder is indicated
in the number of units according to the basic resolution of the encoder. The value is
stored in write-protected memory, which can be changed only by the encoder
manufacturer. This value has practically no importance for the user.
DDLM_Slave_Diag
Octet
36
37
38
39
Bits
31 – 24
23 – 16
15 – 8
7 – 0
Data
231 – 224
223 – 216
215 – 28
27– 20
Offset Value of the Encoder Manufacturer
Table 35 Diagnostics, offset value of the encoder manufacturer
5.6.16 Scaling Parameters Settings
The scaling parameters are set in the DDLM_Set_Prm function. The parameters are
stored in the octets 40 to 47 of the diagnostic data. The "Measuring units per
revolution" and "Total measuring range in measuring units" parameters define the
selected resolution of the encoders. The status bit of the scaling function in the
operating status (octet 9 of the diagnostic data) indicates whether the scaling function
is enabled.