USER'S MANUAL FP-2b

Janvier 2001

Page 13

5  MAINTENANCE

5.1  Shipment instructions

Instruments  sent  to  repair  or  to  calibrate,  inside  or  outside  the

warranty period, must enclose the following information: Company

name,  name  of  the  person  to  contact  if  necessary,  address,

telephone  number,  purchase  receipt  (in  the  case  of  warranty

period),  and  problem  description.  Freights  will  be  charged  to  the

client.

5.2  Battery replacement

-

 

When  the  symbol 

  appears  on  the  LCD  display,  it

indicates that the battery should be replaced.

-

 

After  disconnecting  the  test  leads  and  setting  the  rotary

selector  at  the 

OFF

  position,  remove  screws  on  the  back

cover and open the case.

-

 

Replace the exhausted battery with a new one.

-

 

Set the cover again and fix the rear screws.

Summary of Contents for FP-2b

Page 1: ...DIGITAL MULTIMETER FP 2b USER S MANUAL 99 Washington Street Melrose MA 02176 800 517 8431 TestEquipmentDepot com...

Page 2: ......

Page 3: ...inly SAFETY RULES paragraph The symbol on the equipment means SEE INSTRUCTION MANUAL In this manual may also appear as a Caution or Warning symbol WARNING AND CAUTION statements may appear in this man...

Page 4: ......

Page 5: ...SPECIFICATIONS 3 3 SAFETY RULES 9 3 1 General safety rules 9 3 2 Specific safety rules 10 4 OPERATING INSTRUCTIONS 11 5 MAINTENANCE 13 5 1 Shipment instructions 13 5 2 Battery replacement 13 5 3 Repla...

Page 6: ......

Page 7: ...to 100 V Automatic overrange indication by means of 1 on the display Automatic polarity indication on the DC range Resistance measurement from 0 1 up to 200 M Capacity measurement from 1 pF up to 20 F...

Page 8: ...USER S MANUAL FP 2b Page 2 Janvier 2001...

Page 9: ...to 40 C Storage temperature 10 C to 50 C Power supply 9V battery 6F22 Dimensions W 88 x H 170 x D 38 mm Weight 340 g battery included Included accessories Test leads PP 008 Case DC 231 NOTE Accuracy...

Page 10: ...ncy range 40 Hz to 1000 Hz Overload protection 700 V AC sine wave signal Indication Average calibrated in RMS of sine wave DC Current Range Resolution Accuracy 2 mA 1 A 20 mA 10 A 0 8 reading 1 dig 20...

Page 11: ...rrent 10 A in continuous mode 20 A 15 s maximum Response Average calibrated in rms of sine wave Resistance Range Resolution Accuracy 200 0 1 0 8 reading 3 dig 2 k 1 20 k 10 200 k 100 2 M 1 k 0 8 readi...

Page 12: ...nF 2 F 10 nF 4 reading 3 dig Frequency test Range Resolution Accuracy 2 kHz 1 Hz 2 reading 5 dig 20 kHz 10 Hz 1 5 reading 5 dig Overload protection 250 V DC AC rms Sensitivity 200 mV AC rms Maximum in...

Page 13: ...USER S MANUAL FP 2b Janvier 2001 Page 7 Continuity test Range Threshold 50 Overload protection 250 V DC AC rms Transistors test Range hFE Readout hFE DC gain Base current 10 A Vce voltage 3 2 V...

Page 14: ...USER S MANUAL FP 2b Page 8 Janvier 2001...

Page 15: ...to ensure safety Test leads Use this instrument under the specified environmental conditions The user is only authorized to carry out the following maintenance operations Battery replacement Fuse rep...

Page 16: ...be sure that points under test do not exceed the maximum voltage Keep measuring leads in good condition Between measurements higher than 10 A which extend near to the maximum time specified keep wait...

Page 17: ...est leads to the measurement points For currents higher than 200 mA connect the test leads to the 10 A red and COM black jacks Measuring resistance Connect the black test lead to the COM jack and the...

Page 18: ...s of the circuit under testing If continuity exists i e resistance less than about 50 built in buzzer will sound Testing transistor Set the rotary selector at hFE position and verify if transistor is...

Page 19: ...lephone number purchase receipt in the case of warranty period and problem description Freights will be charged to the client 5 2 Battery replacement When the symbol appears on the LCD display it indi...

Page 20: ...e it with a new one of the following characteristics 200 mA F 250 V Place and fix the lid again 5 4 Cleaning recommendations The cover should be cleaned by means of a light solution of detergent and w...

Page 21: ...0 IG2564...