USER'S MANUAL FP-2b
Janvier 2001
Page 3
2 SPECIFICATIONS
Readout
LCD 3 ½ digits
Reading update
2-3 measurements / s
Overrange indication
"1" displayed only
Polarity indication
"-" displayed automatically for
negative polarity.
Low battery indication
appears on the display
Operating temperature
5 to 40
°
C
Storage temperature
-10
°
C to 50
°
C
Power supply
9V battery (6F22)
Dimensions
W. 88 x H. 170 x D. 38 mm.
Weight
340 g. (battery included)
Included accessories
Test leads
PP-008
Case
DC-231
NOTE:
Accuracy is specified at 23
±
5
°
C and relative humidity
under 75%.
DC Voltage
Range
Resolution
Accuracy
200 mV
100
µ
V
2 V
1 mV
20 V
10 mV
200 V
100 mV
±
(0.5% reading
+ 1 dig.)
1000 V
1 V
±
(0.8% r 2 dig.)
Input impedance:
10 M
Ω
Overload protection:
250 V rms in the 200 mV range
1000 V DC in the other ranges