EXAMPLE B.- Measuring the power of a digital signal
Take the following figure as an example, showing a digital signal:
Measurement conditions:
Central frequency:
650.00 MHz
RF attenuation:
0 dB
Mode: DIGITAL
Figure 9.- Example of digital channel power measurement.
Using the CH POWER scale [32] (blue background) you can read a power
of 45 dBµV.
From the correction chart in figure 8, you obtain the Correction Factor to be
applied to the frequency you are working with (650.00 MHz) which is 0 dB. Therefore,
the actual power of the channel is:
Power [dBµV] = Reading (45 dBµV) + Atten. (0 dB) + Correct. Factor (0 dB) = 45 dBµV
To obtain further information on the digital signal measurement method refer to
Appendix A Digital TV signals.
Summary of Contents for MC-377+
Page 2: ......
Page 4: ......
Page 5: ...SUMARIO CONTENTS SOMMAIRE Manual espa ol English manual Manuel fran ais Fran ais English...
Page 6: ......
Page 8: ...MANUAL DE INSTRUCCIONES MC 377...
Page 14: ...MANUAL DE INSTRUCCIONES MC 377 P gina 6 Mayo 2001...
Page 18: ...MANUAL DE INSTRUCCIONES MC 377 P gina 10 Mayo 2001...
Page 22: ...MANUAL DE INSTRUCCIONES MC 377 P gina 14 Mayo 2001...
Page 42: ...MANUAL DE INSTRUCCIONES MC 377 P gina 34 Mayo 2001...
Page 46: ...INSTRUCTION MANUAL MC 377...
Page 52: ...INSTRUCTION MANUAL MC 377 Page 6 May 2001...
Page 56: ...INSTRUCTION MANUAL MC 377 Page 10 May 2001...
Page 60: ...INSTRUCTION MANUAL MC 377 Page 14 May 2001...
Page 80: ...INSTRUCTION MANUAL MC 377 Page 34 May 2001...
Page 84: ...MANUEL D UTILISATION MC 377...
Page 90: ...MANUEL D UTILISATION MC 377 Page 6 Mai 2001...
Page 94: ...MANUEL D UTILISATION MC 377 Page 10 Mai 2001...
Page 98: ...MANUEL D UTILISATION MC 377 Page 14 Mai 2001...
Page 118: ...MANUEL D UTILISATION MC 377 Page 34 Mai 2001...