USER’S MANUAL. OS-782
02-2010
Page 24
The following Table describes the
Vertical Channel
menu:
Function
menu
Setting
Description
Coupling
AC
DC
Ground
The
DC
component in the input signal is blocked.
The
AC
and dc components of the input signal are
allowed.
Channel
Close
Open
Close the channel.
Open a channel.
Probe
1x
10x
100x
1000x
Select one according the probe attenuation factor to
ensure a correct vertical scale reading.
Invert
Close
Open
Waveform is displayed normally.
Open the
Invert
function of the waveform setting.
4.11.1 Setting the Channel Coupling
With
CH1
taken for example, the measured signal is a sine wave signal
containing a dc offset. Press
F1 Coupling
first and
then
AC
to make an ac coupling
setting. The dc component contained in the tested signal is blocked.
Press
F1 Coupling
first and
then
DC
to make a dc coupling setting. Both dc and
ac components contained in the tested signal are permitted. The waveform is displayed
as the following figure 13, figure 14.
Figure 13.-
AC coupling.