USER’S MANUAL. OS-782
Page 27
02-2010
En
g
lis
h
6.
Again, press the
OSD OPTION
key twice and the screen shows the following:
◄
/
►
Time
V
/
W
- CHM Zero
7. Press the “
OSC
V
” or “
OSC
W
” key to adjust the position of the waveform
Math. Now, look at the display and you will find a screen that looks like the
following Figure 15.
Figure 15.-
Waveform mathematics.
4.13
Setting the Trigger System. TRIGGER Function.
The
Trigger
defines the time when the acquisition of data and display of
waveform start. If it is set correctly, the trigger can turn an unstable display into a
significant waveform.
When starting the acquisition of data, the oscilloscope collects sufficient data to
draw the waveform at the left side of the triggering point. With waiting for the triggering
condition, the oscilloscope is gathering data continuously. After a trigger is detected, the
oscilloscope gathers enough data continuously to draw the waveform at the right side of
the triggering point.
To make a trigger mode setting, do the following:
1. Press
the
MENU
key and the function menu appears at the right of the
screen.
2. Press
the
“
MENU
V
” or “
MENU
W
” key to select
TRIG MODE
and five items
selectable are displayed at the bottom of the screen.