The
[Simulation] menu item enables you to activate all outputs for
test purposes during commissioning. A simulated output remains
activated until you quite the
[Simulation] menu item. It is also pos‐
sible to prime a peristaltic pump, for example, with
[Simulation]
mode. A
Simulation
Relay 1
A0983
9.2.1
mA output 1
mA output 1
Pump 4
Pump 3
Pump 2
Pump 1
Alarm relay
Relay 2
Off
Off
Off
Off
On
Off
Off
Off
On
Off
Fig. 31: Displaying simulation
8.3 Displaying
[Device Information]
Continuous display ➨ ➨ or
[Diagnostics] ➨
[Diagnostics] ➨ or [Device information]
[Device information]
Device info
Identity Code
A1164
9.3.3
Softw. version: 01.00.00.00
Srnr:
15082008
Module rev. 0100
Operating temperature
35.5 °C
Softw. version: 01.00.00.00
Expansion module
Fig. 32: Device Information
8.4 Error messages and warning alerts
Tab. 6: Error messages
Error
Error message text
Cause
Remedy
88
The connection to the
extension module is
faulty
The connection cable has
slipped out of the socket
Check the connection cable and
tighten
Connection problems between
the main and extension unit
Return to the factory for checking
01
The mV input voltage
is too low
Coaxial cable connection dis‐
connected
Check that the coaxial cable connec‐
tion is fitted correctly and re-connect
[Diagnostics]