10.2 Calibration
The calibration menu is for calibration of the following fields:
n
Calibration Powder
n
Calibration Liquid
n
Calibration Flow monitor
n
Calibration Water
Before the first switching on of the system in automatic mode, you must
calibrate the dry feeder and the liquid concentrate pump. The system must
be switched off during calibration. Unit
[Stop]
. If you have not yet cali‐
brated the system and you switch the system to
[AUTO]
mode, then the
following message appears on the display
[Fault: Dry feeder not calibrated]
of Fault:
[Liquid concentrate pump not calibrated]
and the system switches to
[FAULT]
mode.
10.2.1 Calibration Powder
Mode
Menu
Archive
Reset
Calibration Powder
Unit Stop
Water
Powder Liquid
Flow mon.
Calibration time:
Elapsed time:
Enter weight:
Dosing capacity:
A0793
Fig. 29: Dry product calibration
Parameter
Factory setting
Adjustment range
Enter calibration time
60 s
0 - 999 s
Enter weight
-
99999.9
Interrupting the calibration
You can end the calibration at any time by pressing the
[STOP] key.
Carry out calibration
Operating Menu [F2]
51