Calibrating the peristaltic pump
Every 6 months and/or after a hose change, the pumping rate of the peristaltic pump must be checked and, when necessary,
adjusted. For this purpose, place the sample scales with a beaker into the magnetic stirrer and switch on. If necessary, press the
“TARE” button in order to reset the value to 0.00g.
Then select the calibration menu for the peristaltic pump. The test period of two minutes is represented by a progress bar on the
display. On completion of this process, the pumping rate of the pump is shown on the display of the scales and can be edited using
the “+” and “-“ buttons in the following menu and subsequently confirmed by clicking <OK>.
8
Manual stirring mode
The “Magnetic stirrer” menu option enables manual starting of the stirrer in order to measure the pH, for example. The speed can be
increased and reduced using the +/- buttons of the controller. Press the <ESC> button in order to switch the stirrer off again. The
operating time is limited to 20 minutes. The display shows the remaining operating time.
9
Maintenance, repair, customer service
The manufacturer is solely responsible for the original safety characteristics and features of the device. Any guarantee becomes
void if the device is modified in any manner that is neither performed nor approved by the manufacturer.
Certain activities – such as replacing hardware components or internal settings – may only be carried out by qualified personnel.
The analyzer should be checked and calibrated by the manufacturer or qualified technical personnel every two years. Defective
devices must be taken out of operation and returned to the manufacturer.
Pronova Analysentechnik GmbH & Co. KG
Groninger Straße 25
13347 Berlin, Germany
Telephone
+49 (0)30 / 455 085 –0
Fax
+49 (0)30 / 455 085 –90
Web
Stirrer
active 18:06
Speed +/-
Measuring
Calibration
Stirrer
A/D Values
OK
Calibration
Peristaltic
Pump with
< OK >
Mass 5.0 g
Continue
with
< OK >
Calibration
active
__________
OK
Calibration
pH Electrode
Perist. Pump
OK
Summary of Contents for FOS/TAC 2000
Page 2: ......
Page 4: ......
Page 21: ...FOS TAC 2000 APPENDIX 812 04 2015 EN FOS TAC 2000 User Manual 21 11 Appendix Appendix ...
Page 22: ......