Press MODE after treatment mode is set.

1. To choose Program (P1~P8), use Up + (to increase) or the button Down – (to 

decrease).

2. Program P9, it is self –programmable mode.  To set up desirable  pulse width 

between 50~260uS and frequency at  2~60Hz (for TENS) and 7~70Hz (for EMS) : 

2.1. Press Mode button, the preset (default) Pulse Width is then flashing. 

Press the button Up + (to increase) or the button Down – (to decrease) for choice 

of the selected Pulse Width . 

 

2.2. Press Mode button again,  the preset (default) Pulse Rate is then flashing.  

Press the button Up + (to increase) or the button Down – (to decrease) for

choice of the selected Pulse frequency.

 

3.  Press MODE again to save your selection. The treatment program you selected 

will appear on the display the next time you turn the device on.

 

 

 

 

 

 

Selecting the Treatment Time

1.  Press MODE. The preset (default) treatment time will flash on the display.

2.  To increase or decrease the treatment time, press the button ON + (to increase) 

or the button OFF – (to decrease) repeatedly until the desired duration appears 

on the display.

3.  Press MODE again to save your selection. The treatment time you selected will 

appear on the display the next time you turn the device on.

Note:  

If you change programs during the course of a therapy session, 

             the treatment time will not reset unless you manually reset it by performing      

             the steps described above.

Selecting the Therapy Intensity Level

This device offer a maximum of 25 intensity levels.    

The intensity of the electrical current determines the number of working fibers in 

the stimulated muscles. The lower the current intensity the lower the number of 

working fibers in the muscle. The higher the current intensity the greater the num-

ber of working fibers in the muscle. 

If using the device for help with temporary relief of pain associated with sore and 

ching muscles then you will find that setting the current intensity to your own com-

fortable and pleasing level will give you much satisfaction. This level is different for 

each user so adjusts slowly and accordingly. 

If you desire to stimulate healthy muscles (EMS)  in order to improve and facilitate 

muscle performance then you want to achieve a significant number of working  

fibers. You require a minimum intensity (approx. 30mA) to accomplish this . 

This can be achieved relatively quickly (2 to 3 sessions) by progressively increasing 

the intensity during the session. Once this threshold is reached, continue to pro-

gressively increase the current intensity making the session more effective. 

 

      

  

Summary of Contents for SUPER DOU plus

Page 1: ...OPERATING INSTRUCTION SUPER DUO TENS EMS plus 1639...

Page 2: ...e carefully and keep them for lateruse besuretomakethemaccessibletootherusersand observe the information they contain Signs and symbols WARNING Warning instruction indicating a risk of injury or damag...

Page 3: ...e electrical impulses that are sent into the tissue influence the transmission of stimulation into nerves nerve centres and muscle groups in the application area Electrostimulation usually only has an...

Page 4: ...ormance e g after a marathon Thanks to integrated massage technology TENS EMS SuperDuoPlus units are also able to relieve muscle tension and combat fatigue with a program based on the sensation and ef...

Page 5: ...rs If you are receiving medical treatment In the event of complaints linked to stimulation treatment If you suffer from persistently irritated skin due to long term stimulation at the same electrode s...

Page 6: ...dhesive capability of the adhesive electrodes decreases please replace them immediately Only use the device again with new adhesive electrodes Otherwise the unequal adhesion of the adhesive electrodes...

Page 7: ...s to the stimulator use the following steps to prepare your skin at the electrode placement sites 1 Determine the placement sites for the electrodes 2 Wash the area with mild soap and water do not use...

Page 8: ...Always ensure the limb is secured to prevent movement 2 Do not turn the unit on until all electrodes and lead wires are properly attached Placement of the Electrode Pads forTENS Treatment of Pain Note...

Page 9: ...gra dually You may use any of the modes for EMS SHOULDERS Place one half of the pad on the front of your shoulder and the otheron the side BACK OF THIGHS Place each pair of pads horizontally across y...

Page 10: ...ation Program Pulse width uS Pulse rate Hz Waveform Type P1 P2 P3 P4 P5 P6 P7 P8 P9 260 260 260 260 156 260 156 260 260 156 50 260 uS 15 60 60 2 60 60 7 60 60 Constant Burst Constant Dense Disperese M...

Page 11: ...time you turn the device on Note If you change programs during the course of a therapy session the treatment time will not reset unless you manually reset it by performing the steps described above S...

Page 12: ...s 5 10 15 20 25 30 35 40 45 50 55 and 60 min utes Time will countdown on the display in 1 minute increments for the duration of your session The device turns off automatically when the therapy time ha...

Page 13: ...d electrodes Do not pull on the cables but on the connectors attached to the ends of the cables Store the stimulator with the cables in a clean dry place Electrode The electrode pads are disposable an...

Page 14: ...completely flat batteries should be disposed of through specially designated collection boxes recycling points or electronics retailers You are legally required to dispose of the batteries For environ...

Page 15: ...re he can send the device free of cost and what documentation is required Filing a warranty claim can only come into consideration if the buyer can submit a copy of the invoice proof of purchase and t...

Page 16: ...Well Life Healthcare Ltd 6F No 168 Lide St Jhonghe District New Taipei City 23512 Taiwan EXCLUSIVELY IMPORTED AND DISTRIBUTED IN EUROPE BY EUROMEDICS GmbH Hauptstra e 169 D 59846 Sundern Germany info...

Reviews: