26
TENS+EMS SUPER DUO STIMULATOR
TECHNICAL SPECIFICATIONS
Channel: Dual, isolated between channels.
Output Voltage: 0~40V peak into 500Ω load each channel.
Pulse Amplitude: Adjustable 0 – 80mA.
Pulse Rate: As pre-programming operation mode.
Pulse Width: As pre-programming operation mode.
Software ramp up feature: Pulse width ramp up when change mode.
Timer : 5~60 min. selectable.
LCD: Show modes, pulse rate, pulse width, timer, CH1/CH2, intensity level.
Wave Form: Symmetrical Bi-Phasic square pulse.
Max Charge per Pulse: 20.8 micro-coulombs maximum.
** All electrical specifi cations are ±20%
Operating Conditions: + 50°F (10
°C
) to +104° (40
°C
), 40-90% max. Relative humidity
Transport and Storage Temperature: +14°F (-10
°C
) to +140° (60
°C
), 30-95% max. Relative
humidity
Weight: 120 g (battery included)
Dim. : 122 x 66 x 30 mm
Power Source: 3 x AAA / 4.5 Volt batteries
Technical Specifi cations
This symbols means “ Serial number “
This symbols means “ Attention, consult the accompanying
documents”
This symbols means “ Manufacturer “
This symbol means type BF equipment; this device off ers
protection against electrical shock by standard compliance to
leakage currents of electrode pad.
This device shall be disposed in accordance with national laws
after their useful lives
This symbol means “used before”, represent as “YYYY-MM”
(for year and month).
(i) There are a number of technical symbols on your unit explained as follows:
(ii) there is a label on the package of electrode explained as follows: