OPERATION
1. Place scale on a horizontal flat surface, press
[ON/OFF] key to turn on the scale.
2. Wait until "0.0" is displayed.
3. Put the object(s) on the weighing platform
4. Using the [MODE] key, you can change the
weighing units.
TARE
1. Turn on scale as described above.
2. Place the "tare item" on the platform.
3. Press [TARE] and wait until "0.0" is displayed.
4. Add the "net-weight-item”.
CALIBRATION
1. Place the scale on a stable surface and turn
on the scale.
2. Press and hold [MODE] button and release
when the display shows “CAL”.
3. Press [MODE] again, the LCD display will
show the required calibration weight.
4. Gently place the required weight on the
scale and and wait 3 seconds.
5. The display will show “PASS” when calibra-
tion is complete and return to the normal
weighing mode.
SPECIFICATIONS
Tare Range
Full capacity
Auto Off
60 seconds off
Op. Temp
15 to 30 ˚C
Power
2 x AAA
Scale Dim.
110 x 70 x 15mm
Tray
68 mm x 47 mm
LCS
100 / 500
Capacity
100g / 500g
Readability
0.01g / 0.1g
Modes
g, oz, ct, gn
DISPLAY ERROR MESSAGES
[Lo]
Change the batteries.
[O-LD]
Overload
[UnST]
Use the scale in a more
stable position.
[OUT2]
Calibration Required
Summary of Contents for LCS100
Page 1: ...USER MANUAL ...