background image

OPERATION

1. Place scale on a horizontal flat surface, press 
    [ON/OFF] key to turn on the scale. 
2. Wait until "0.0"  is displayed. 
3. Put the object(s) on the weighing platform 
4. Using the [MODE] key, you can change the 

weighing units.

TARE

1. Turn on scale as described above.
2. Place the "tare item" on the platform.
3. Press [TARE] and wait until "0.0" is displayed.
4. Add the "net-weight-item”.

CALIBRATION

1. Place the scale on a stable surface and turn 

on the scale.

2. Press and hold [MODE] button and release 

when the  display shows “CAL”.

3. Press [MODE] again, the LCD display will 

show  the required calibration weight.

4. Gently place the required weight  on the 

scale and  and  wait 3 seconds.

5. The display will show “PASS” when calibra-

tion is complete and return to the normal 
weighing mode.

SPECIFICATIONS
Tare Range

  Full capacity

Auto Off

 

60 seconds off

Op. Temp

 

15 to 30 ˚C

Power

 

2 x AAA

Scale Dim.

 

110 x 70 x 15mm

Tray

 

68 mm x 47 mm

LCS 

100 / 500

Capacity

 

100g / 500g

Readability

  0.01g / 0.1g

Modes

 

g, oz, ct, gn

DISPLAY ERROR MESSAGES
[Lo]                     

Change the batteries. 

[O-LD]                 

Overload

[UnST]               

Use the scale in a more

                              stable position. 

[OUT2]               

Calibration Required