5
“OFF” range position, this provides direct short across meter
movement for minimum needle bounce when transporting
meter.
9. If you should accidentally apply excessive voltage or current
on a certain range, disconnect the leads from the circuit as
quickly as possible, check instrument operation on that range
by applying peoper input. If the meter does not operate
properly, check fuse. If it is blown replace it. (See fuse
replacement.)
OPERATION PROCEDURES
DC Voltage Measurement
WARNING: USE EXTREME CARE WHEN MAKING
MEASUREMENTS FOR HIGH VOLTAGE. DO NOT TOUCH
TERMINAL OR PROBE ENDS.
1. Set the selector switch to the appropriate DCV range to be
used.
2. Connect the BLACK test lead to the “-COM” jack and the RED
test lead to the “+” jack.
3. If you know the polarity of the circuit to be tested, connect the
black probe to the negative side.
4. If you don’t know the polarity, connect the probes to opposite
sides of the circuit and watch the pointer. If it goes to the left,
reverse the probes. The RED probe will be connected to the
positive.
5. Check the needle position and the get the reading on V.A scale.
Null DCV (Central Zero) Measurement
At these two ranges, it can automatically judge the polarity of
circuit as the pointer can move to the center line and become a
Null meter.
1. Set the selector switch to the DCV ±5V or 25V range.
2. Connect the BLACK test lead to the “-COM” jack and the RED
test lead to the Red “+” jack.
Summary of Contents for mt-2017
Page 18: ...17 Diode 10K 1K 100 10 1 A 1 A 10 A 100 mA 1 mA 10 1 2 COM 3 Rx10 4 1 6 2 9V 3 4...
Page 21: ...20 DC10A 10A 1 1 5V 90 90 2 9V...
Page 22: ...21 3 90 0 5A 250V 10A 250V 0 5A 250V 10A 250V 5 20mm UL 90 1 1 5V 1 5V 9V...
Page 26: ...25 Diode 10K 1K 100 10 1 A 1 A 10 A 100 mA 1 mA 10 1 2 COM 5 Rx10 6 1 6 2 9V 3 4...
Page 29: ...28 DC10A 10A 1 1 5V 90 90 2 9V...
Page 30: ...29 3 90 0 5A 250V 10A 250V 0 5A 250V 10A 250V 5 20mm UL 90 1 1 5V 1 5V 9V...
Page 31: ...30 MT 2017 C 12 CD 0755 83692415 83692986 83246594 83247554 0755 83692143...