E.
Repeat the above procedure for additional measurements.
iX. Suggestion “A” for material Static decay measurements
Overview:
The dual phono plug (input) cable assembly is connected to the analog output of the PFM-711A Field Me
-
ter and the Analog Input of the PDT-740B Decay Timer. The Charge Plate Assembly (CPM-720) is installed
on the Field Meter. One end of a high resistance lead is connected to the CPM-720 Plate and the other
end connected to a Prostat PGB-745 Electrode. The PCS-730 Charging Source is used to apply greater
than ±1,000 volts to the CPM, which is shared with the PGB-745 Electrode. To measure material’s decay
time performance, refer to the figures below and proceed as described.