PGA-710 Autoanalysis System – How To Measure Guide
22
IV. Instrument Test Setup for Decay
The instrument setup for static decay measurement of ionizers is similar to body voltage
measurement.
1. Mount the Prostat CPM-720 Charge Plate to the PFM-711A Field Meter and connect the
field meter to ground. (See the instrument’s operations manual for detailed instructions.)
2. Connect the analog output of the field meter to the input of the PGA-710 Autoanalyzer
using the supplied lead.
3. To prevent system damage to the PRO
STAT
®
PGA-710 Autoanalysis System and
voiding your warranty, input to the Autoanalyzer
may not exceed ±2.0 volts
without
specific adjustments and recommendations from PROSTAT. Confirm proper non-
PROSTAT instrument output voltages before connecting to, and energizing the system
4. Use the Prostat PCS-730 ±1 kV Charging Source for applying voltage to the charge plate
for decay testing. Be sure the PCS-730 is grounded for proper operation.
CAUTION
To prevent system damage to the PRO
STAT
®
PGA-710 Autoanalysis System and
voiding your warranty, input to the Autoanalyzer
may not exceed ±2.0 volts
without
specific adjustments and recommendations from PROSTAT. Confirm proper non-
PROSTAT instrument output voltages before connecting to, and energizing the
t
Figure 37: Ground Charging Source for Proper Operation