PRF-911 Concentric Ring
13
g. Select Ohms Mode by pressing “Mode” button once.
1. Push Test Button
2. Resistance Should be < 10
Ω
(1) If proper connection is made through the BLACK receptacle to
the outer ring, and the outer electrode is in contact with the
metal test bed, the PRS-801 or PRS-812 will indicate low resistance is < 10
Ω
(2) If proper continuity indication confirms positive contact of the
outer electrode, the fixture check is completed.
(3) If no continuity indication appears, clean the electrode surface with 70%
solution of IPA , test the PRS-801 or PRS-812 per its operating instructions
and repeat the procedure. (i.e. Install the supplied shunt and test the low
range. Check the cable & connections as well.)
IV. General
Operation
A. Surface Resistance Measurement per ESD STM 11.11-2001. Note that this procedure
replaces surface resistivity measurements of electrostatic discharge control materials
previously conducted using ASTM D-257. The following procedure is used to measure
surface resistance in ohms, and convert it to ohms/square if necessary.
1. Items necessary for this procedure
a. PRF-911 Concentric Ring Fixture
b. BNC/Dual Banana Adapter
c. PTB-920 Dual Test Bed
d. PRS-801 or PRS-812 Resistance Instrument and test leads
e.One PROSTAT five pound NFPA electrode
f. One GREEN auxiliary ground reference lead or PRS-801 High Resistance Lead
harness if using PRS-801 Instrument. NOTE: Connection to instrument ground
reference is recommended for measurements of 1.0x10
Ω
and higher
g. Clear worksurface
2. Specimen Preparation for Surface Resistance measurement
a. This procedure is designed specifically for measuring planar (flat)
surfaces having relatively smooth surfaces.
b. In the laboratory environment, sample preparation usually includes
cutting a 3x5 inch specimen of the sample material to be tested. This