PRF-911 Concentric Ring
14
insures standard specimen size and minimizes fringe voltages that may
effect the surface resistance measurement.
c. In the auditing process, one often must obtain measurement of a large
piece of material without cutting sample into a proper size specimen.
This type of measurement is considered "indicative"; however, it is
sufficient for obtaining a reasonable indication of the material's surface
resistance PROVIDING only one layer of the material is measured as
described below.
3. PRF-911 Concentric Ring Fixture Preparation
a. Install the BNC/Dual Banana adapter on the PRF-911 fixture
b. Check proper battery level of the PRS-801 or PRS-812 Resistance Instrument
and connect the meter leads to the PRF-911 Fixture as follows:
(1) Install the instrument negative (-) lead of the PRS-801 or PRS-812
Resistance Instrument to the RED BNC adapter receptacle connected to the
center INNER RING electrode (D1).
(2) Install the positive (+) lead to the BLACK receptacle connected to the OUTER
RING electrode (D2).
(3) To enhance the accuracy of the measurement, connect a
GREEN auxiliary ground lead to the BNC adapter fixture ground
receptacle, and the other end of the lead to the instrument’s ground
reference, if so equipped.
c. Place the PTB-920 Dual Test Bed on a worksurface with the insulated
acrylic surface facing UP .
Figure 10: PRF-911 Fixture Preparation
Instrument Ground Reference
Positive ( + ) Test
Voltage Power Lead
Negative ( - ) Sensing Lead