background image

9

Rev. A  / November 2005

PRF-914B High Resistance Miniature Concentric Ring Fixture Set

Figure 7: Connect Instrument Adapter to 

PRS-801. Plug (+) Source banana into Posi

-

tive instrument terminal. Plug (_) Sense 

banana into (-) Negative terminal

Figure 8: Connect Instrument Adapter 

ground to instrument’s ground reference

B.  Confirming Proper BNC Connections, Continuity & High Resistance Tests 

 

The following confirms proper connections by checking continuity of the micro probe against a 

metal plate, then confirms its ability to measure high resistance.

1.  To confirm general setup and function of the PRF-914B, place the electrodes against a clean 

metal surface. For example, the plated metal side of the Prostat PTB-920, Dual Surface Test 

Bed. 

a.  Hold the PRF-914B vertically, and apply pressure to slightly compress the electrodes, mak

-

ing positive contact with the metal surface.

b.  Activate the wide range, resistance instrument to obtain a measurement.

 

 

In the case where the Prostat PRS-801 is the measurement instrument, it should   

 

 

 

measure approximately 1.0 ohm, or less.

2.  Repeat the above procedure using the clean insulated surface (Black, labeled side) of the PTB-

920, or an insulated acrylic plate.

a.  Hold the PRF-914B vertically, and apply pressure to compress the electrodes, making posi

-

tive contact with the insulated surface.

b.  Activate the wide range, resistance instrument to obtain a measurement.

 

 

In the case where the Prostat PRS-801 is the measurement instrument, it should  measure  

 

 

1.0x10

14

 ohms, or greater.

C.  PRF-914B Verification Using the PRV-913B

1.  Connect the BNC/banana adapter to the wide range, resistance measurement instrument. 

Flange side of dual banana should be connected to the Positive (+) power terminal. This ap

-

plies test voltage to the 10 outer ring electrodes.

Summary of Contents for PRF-914B

Page 1: ...User Manual PRF 914B MINIATURE CONCENTRIC RING FIXTURE ...

Page 2: ......

Page 3: ...nce 11 V Warranty Information 13 General Specifications 14 Table of Contents Copyright 2012 by Prostat Corporation All rights reserved Printed in the United States of America No part of this manual may be used or reproduced in any manner whatsoever without written permission For information contact Prostat Corporation 1072 Tower Lane Bensenville IL 60106 USA Prostat is the registered trademark of ...

Page 4: ...tely 1 10th the size of the S11 11 fixture Based on its size the PRF 914B is often referred to as a micro probe 1 The resistance measurement range of the PRF 914B fixture is 0 9 ohms at 10 volts to 1 0E 14 1 0x1014 ohms at 100 volts 2 The miniature concentric fixture consists of a spring loaded 0 10 inch 2 54 mm diameter center electrode surrounded by 10 each 0 06 inch 1 5 mm diameter contact elec...

Page 5: ...2B PRF 914B and PRF 922B Microprobe Resistance Fixtures It consists of a high quality black anodized housing and circuit board incor porating an array of ten 10 each 10 megohm precision 2 resistors in parallel connecting the center pad to each of the ten peripheral trace pads it provides a specific reference resis tance of 1 0x106 ohms 5 on the concentric ring side On the other side it consists of...

Page 6: ...formance micro probed designed for use with a maximum input voltage of 100 volts As such it is capable of delivering an annoying shock to any person touch ing the spring loaded contacts when they are energized a If used with the Prostat PRS 801 Resistance System the current capability of the micro probe instrument combination is limited to a very low typically harmless level However a distinct haz...

Page 7: ...tampering or attempted re pair will absolutely void product warranty and completely absolve Prostat Corporation of any responsibility liability or other whatsoever 4 Do Not Touch Electrode Surfaces Electrodes will become contaminated with skin oils and salts and may become damaged or rendered inaccurate 5 Do Not Use Or Store PRF 914B Or PRV 913B In Damp Environments Always store devices with prote...

Page 8: ...e connectors Tighten finger tight only 2 Connect the Sense cable s BNC fitting to the instrument Adapter s Sense terminal 3 Connect the Source cable s BNC to the instrument Adapter s Source terminal Figure 6 4 Once the cables are installed and connected to the Instrument Adapter the PRF 914B is ready for connection to a wide range ohmme ter continuity test verification and use 5 Connect Instrument...

Page 9: ... slightly compress the electrodes mak ing positive contact with the metal surface b Activate the wide range resistance instrument to obtain a measurement In the case where the Prostat PRS 801 is the measurement instrument it should measure approximately 1 0 ohm or less 2 Repeat the above procedure using the clean insulated surface Black labeled side of the PTB 920 or an insulated acrylic plate a H...

Page 10: ... and lower it until the spring loaded center electrode makes direct contact with the material under test 3 Apply sufficient pressure on the probe until the center and outer spring loaded electrode are partially compressed while in contact with the test material 4 Select appropriate instrument test voltage and initiate resistance measurement ESD Association S11 11 test voltage guidelines for measur...

Page 11: ...probe individually by grasping it firmly then pulling it straight out of its socket b Inspect each probe for damage then clean with the alcohol solution If a probe is damaged i e bent does not compress smoothly or has deep surface scratches replace it with a new probe of the same size and characteristics Contact Prostat Corporation Customer Service for spare replacement probes c Clean and dry the ...

Page 12: ... Verification Fixture 1 Store the PRV 913B in a clean dry environment 2 Periodically clean and dry the gold fixture contact segments twice with a solution of laboratory grade isopropyl alcohol and laboratory quality swab Figure 13 View of PRF 912B Spring Loaded Test Pins Socket Teflon Mounting Disk Figure 14 Insert Test Probe into its Socket and Press Firmly to fully seat ...

Page 13: ... not limited to contract negligence or strict liability Fulfillment by Prostat of its express warranty obligations described above will be purchaser s exclusive remedy and will be Prostat s and seller s limit of liability for any breach of warranty or otherwise B Shipping of Warranty Returns 1 Obtain a Return Materials Authorization RMA number and shipping address from Prostat customer service Pac...

Page 14: ...meter Size Probe Spring 3 5 pounds 1 6 kg Force Test Probe Total Travel 0 3 inches 7 8mm Connection BNC with outer source and inner sense connections Cable Dual RG 174 coaxial cables equipped with SMA connectors Instrument Adapter for connection to Prostat PRS 801 Resistance System equipped with BNC male terminals to fit cable BNC female fittings Power Not applicable Fixture powered by resistance ...

Page 15: ...15 Rev A November 2005 PRF 914B High Resistance Miniature Concentric Ring Fixture Set NOTES ...

Page 16: ...16 Rev A November 2005 PRF 914B High Resistance Miniature Concentric Ring Fixture Set NOTES ...

Page 17: ...17 Rev A November 2005 PRF 914B High Resistance Miniature Concentric Ring Fixture Set NOTES ...

Page 18: ...18 Rev A November 2005 PRF 914B High Resistance Miniature Concentric Ring Fixture Set NOTES ...

Page 19: ......

Page 20: ...ville IL 60106 630 238 8883 Fax 630 238 9717 1 855 STATIC1 www prostatcorp com Prostat Corporation Specifications are subject to change without notice All Prostat trademarks and trade names are the property of Prostat Corporation All other trademarks and trade names are the property of their respective companies ...

Reviews: