background image

4

Rev. 2  / 3-18-2019

PRF-922B Miniature Two-Point Probe Set

I. Introduc

Ɵ

 on & Descrip

Ɵ

 on

The PRF-922B Miniature Two-Point Resistance Fix-

ture Set accurately measures surface resistance of 
small areas. It consists of a PRF-922B Two-Point Re-
sistance Fixture assembly, shielded cable equipped 
with BNC connectors; probe cover, and a BNC to 
male banana adapter.

The PRF-922B works in conjunc  on with a wide 
range resistance instrument, such as the Prostat 
PRS-801 Resistance System, PRS-812 Resistance 
Meter, or ESD Check ESI-870 Resistance Indicator, 
and an insulated test bed. This instrument and 

fi

 xture combina  on provides direct point-to-point 

resistance measurements in ohms across the sur-
face of small parts and assemblies.

A.  PRF-922B Two-Point Resistance Fixture Descrip-

 on

This precision 

fi

 xture consists of 2 gold plated, 

spring-loaded contact electrodes spaced 0.250 inches (6.35mm) apart, center-to-center. The elec-
tronic quality gold plated contact probes are 0.1-inch (2.54mm) in diameter, and are supplied with 
conduc  ve synthe  c rubber contact boots, which are 0.125-inch (3.18mm) diameter. The conduc-

 ve boots are used to reduce contact resistance between the gold contact probes and materials 

under test in certain applica  ons.

 

1.  The resistance measurement range of the PRF-922B 

fi

 xture is 0.9 ohms at <10 volts to 1.0E+12  

(1.0x10

12

 ) ohms at 100 volts.

 

2.  Contacts are pogo-pin type ATE quality probes made of beryllium Copper, coated with 60 micro 

inch of hard Gold. Spacing allows for measurement of items with surfaces approximately 0.125 
inches wide having a length of 0.32 inches or larger.

3.  The contact assembly’s outer housing incorporates “stops” that insure consistent contact pres-

sure during measurement.

 

4.  Overall size of the PRF-922B is 0.50 inches in diameter by 7.0 inches (178 mm) long. This op  -

mal size and shape make the 

fi

 xture very comfortable and easy to handle. Its outer housing is 

made of black anodized aluminum.

II. Cau

Ɵ

 ons & Warnings

As with any electrical device, use proper electrical precau  ons and measurement prac  ces to avoid 
personnel shock.  Read this manual in its en  rety before a  emp  ng to use these products.

NOTE

This manual displays Cau  ons and Warnings aler  ng the user to hazardous opera  on and ser-
vicing condi  ons.  

CAUTION

 or 

WARNING

 headings throughout this publica  on 

fl

 ag this infor-

ma  on, where appropriate. Follow all Cau  on and Warning instruc  ons at all   mes.

Figure 1: PRF-922B Miniature Two-Point Resistance 
Fixture Measurement Set.

Summary of Contents for PRF-922B

Page 1: ...PRF 922B User Manual MINIATURE TWO POINT PROBE SET ...

Page 2: ......

Page 3: ...esistance to Ground Rtg Measurements 11 VII Handling Maintenance 12 VIII Warranty Informa on 13 General Specifica ons 15 Table of Contents Copyright 2012 by Prostat Corpora on All rights reserved Printed in the United States of America No part of this manual may be used or reproduced in any manner whatsoever without wri en permission For informa on contact Prostat Corpora on 1072 Tower Lane Bensen...

Page 4: ...t resistance between the gold contact probes and materials under test in certain applica ons 1 The resistance measurement range of the PRF 922B fixture is 0 9 ohms at 10 volts to 1 0E 12 1 0x1012 ohms at 100 volts 2 Contacts are pogo pin type ATE quality probes made of beryllium Copper coated with 60 micro inch of hard Gold Spacing allows for measurement of items with surfaces approximately 0 125 ...

Page 5: ...e maximum applied opera ng test voltage of 100 volts d Only qualified instrument repair and test personnel should exceed the 100 volt opera on limit and then do so only under controlled condiƟons using maximum precauƟons against personnel shock e Never under any condi ons exceed 500 volts during fixture test or repair B Other Opera onal Precau ons 1 Do Not Use the PRF 922B Micro Probe if it fails ...

Page 6: ... arc discharge which in turn may cause the igni on of combus ble materials or fumes Do not use exposed energized circuits in flammable areas 7 Do not a empt to measure energized materials items or circuits with the PRF 922B 8 The PRF 922B is a precision fixture to be operated by experienced personnel familiar in the use and handling of devices employing energized power supplies 9 Do Not Drop or ca...

Page 7: ...etal plate then confirms its ability to measure high resistance 1 To confirm general setup and func on of the PRF 922B place the electrodes against a clean metal surface For example the plated metal side of the Prostat PTB 920 Dual Surface Test Bed a Hold the PRF 922B ver cally and apply pressure to slightly compress the electrodes mak ing posi ve contact with the metal surface b Ac vate the wide ...

Page 8: ...ring electrodes 2 Remove PRF 922B Probe Cover Figure 5 3 Posi on PRF 922B ver cally into the PRV 913B Two Point Verifier with its spring loaded pin electrodes making direct contact with the Verifier s gold plated test segments 4 Depending on your resistance instrument select either 10V or 100V test voltage 5 Measure PRF 922B probe resistance while posi oned in the PRV 913B Two Point Verifier Resis...

Page 9: ... voltage guidelines for measuring packaging materials are as follows a For material resistance measurements of less than 1 0x104 ohms use 10 volts b For measurements of 1 0x104 to 1 0x106 ohms use 10 volts c For measurements greater than 1 0x106 ohms use 100 volts NOTE For op mal performance and accuracy use the Prostat PRS 801 Resistance System in its AUTOMATIC Mode either Default Mode 1 Ohms or ...

Page 10: ...onduc ng pin Figure 9 Removing the inacƟve pin helps idenƟfy point of actual measurement by the acƟve pin C Once pin con nuity is determined and connec on confirmed install a conduc ve rubber boot on the ac ve measurement pin as shown below as shown in Figure 5 The boot reduces contact resis tance and enhances measurement repeatability It is required by ANSI ESD STM11 13 V Material Device Measurem...

Page 11: ...may be Groundable depending on its measured resistance characteris cs VI Resistance to Ground Rtg Measurements Rtg measurements are made on installed materials or assemblies while used in the manufacturing pro cess ESD control materials and tools are intended to func on while properly grounded Rtg measurement confirms the material is a actually grounded and b the total resistance of the material w...

Page 12: ...e posi ve test lead is separately connected to ground VII Handling Maintenance A PRF 922B Miniature Two Point Probe Fixture 1 Store the PRF 922B in a clean dry environment with both BNC and Probe covers installed for environmental and mechanical protec on 2 Periodically remove all spring loaded test pins Clean the spring loaded test probes and Teflon moun ng disk with a solu on of laboratory grade...

Page 13: ...stat will replace at its expense such parts that it determines to be defec ve Any defec ve part must be returned to PROSTAT postage prepaid with proof of purchase date Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESS AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warr...

Page 14: ... and shipped as described above and as di rected by Prostat s customer service department 2 The following informa on must be included with the returned product a Descrip on of the problem b Customer s Purchase Order Number Prostat s Materials Authoriza on RMA number c Name telephone number and fax number of individual contact who can provide more informa on regarding the problem and related applic...

Page 15: ...l Travel 0 3 inches 7 8mm Connec on BNC with outer source and inner sense connec ons Cable RG 174 coaxial cable equipped with insulated BNC couplers and BNC Banana adapter for connec on to Prostat PRS 801 Resistance System Power Not applicable Fixture powered by resistance instrument Warranty Prostat Corpora on Limited one year PRV 913B Dual Verifica on Fixture Physical Dimensions 2 0 in x 2 0 in ...

Page 16: ...16 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...

Page 17: ...17 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...

Page 18: ...18 Rev 2 3 18 2019 PRF 922B Miniature Two Point Probe Set NOTES ...

Page 19: ......

Page 20: ...Fax 630 238 9717 1 855 STATIC1 www prostatcorp com Prostat Corpora on Specifica ons are subject to change without no ce All Prostat trademarks and trade names are the property of Prostat Corpora on All other trademarks and trade names are the property of their respec ve companies ...

Reviews: