background image

 

 

-  3  -

 

© 201

https://www.ProTechTrader.com

 

 

 
Technical Specifications 

Accuracies are guaranteed for 1 year, 23°C±5°C, 
less than 80%RH 

DC VOLTAGE 

RANGE RESOLUTION

ACCURACY

200mV

100uV

±(0.5% of rdg + 3D)

2000mV

1mV

20V

10mV

200V

100mV

1000V

1V

±(1.0% of rdg + 5D)

±(0.8% of rdg + 5D)

 

OVERLOAD  PROTECTION:  220V  rms  AC  for 
200mV range and 1000V DC or 750V rms for all 
ranges. 
 

AC VOLTAGE 

RANGE RESOLUTION

ACCURACY

200V

100mV

750V

1V

±(2.0% of rdg +10D)

 

RESPONSE:  Average  responding,  calibrated  in 
rms of a sine wave. 
FREQUENCY RANGE: 45Hz ~ 450Hz   
OVERLOAD PROTECTION:  1000V  DC  or  750V 
rms for all ranges. 

 

 

Summary of Contents for PDM-711

Page 1: ...tps www ProTechTrader com User Guide ProTechTrader LARGE SCREEN DIGITAL MULTIMETER PDM 711 Please read this manual before using the multimeter Latest manual and more can be found on https www ProTechT...

Page 2: ...n DC or 30V rms in AC take additional care due to increased danger of electric shock Use the proper terminals function and range for your measurements Do not use or store the Meter in a high temperatu...

Page 3: ...rdg 3D 2000mV 1mV 20V 10mV 200V 100mV 1000V 1V 1 0 of rdg 5D 0 8 of rdg 5D OVERLOAD PROTECTION 220V rms AC for 200mV range and 1000V DC or 750V rms for all ranges AC VOLTAGE RANGE RESOLUTION ACCURACY...

Page 4: ...AD PROTECTION 500mA 250V fuse 10A range unfused MEASURING VOLTAGE DROP 200mV RESISTANCE RANGE RESOLUTION ACCURACY 200 0 1 1 0 of rdg 10D 2000 1 20K 10 200K 100 2000K 1K 1 0 of rdg 4D MAXIMUM OPEN CIRC...

Page 5: ...sired DC AC VOLTAGE position if the voltage to be measured is not known beforehand set switch to the highest range and reduce it until satisfactory reading is obtained AC 3 Connect test leads to devic...

Page 6: ...to desired DCA position 3 Connect test leads in series with the loaded current to be measured so that current may flow through the meter 4 Read current value on Digital Display Please Note The 10A fun...

Page 7: ...ower before measurement 4 Connect test leads to circuit being measured 5 Read resistance value on Digital Display DIODE MEASUREMENT 1 Red lead to V mA Black lead to COM 2 Set RANGE switch to position...

Page 8: ...lector leads Insert the leads into the proper holes of the hFE Socket on the front panel 3 The meter will display the hFE value while providing the base current 10 A and VCE2 8V CONTINUITY TEST Beep 1...

Page 9: ...instrument is warranted to be free from manufacturing defects in material and workmanship for one year Any instrument found defective within one year may be returned to the factory with transportation...

Page 10: ...electricity Approximately 4mA Product Size 129 x 70 x 29mm Product net weight 129g including battery Battery and Fuse Replacement Fuse will rarely need replacement Blown fuses are likely a result of o...

Reviews: