page 59
cos105000-2016-00 en
Option Description
Comment / Display
OP29 Standard acceleration and deceleration level for
RS 232 interface
The selected acceleration and deceleration level is valid for control
and operation via RS232 interface. This option is very helpful if the
peripheral equipment (e.g. ECG, ergospirometry, PC) does not offer a
menu for acceleration and deceleration levels.
flashing, reports: 1 ... 5, (standard: 1) for the acceleration and
deceleration level for all speed commands via RS 232. The maximum
adjustable value depends on the setting of option 28.
Note: If the peripheral equipment sends an acceleration and
deceleration command via the h/p/cosmos coscom protocol, the
selected level in option 29 or in the corresponding command of the
h/p/cosmos coscom protocol.
OP40 Locking and unlocking of the treadmill
OFF = After switching on, the treadmill is completely locked / not
accessible. To unlock the treadmill, press the b, - and START
simultaneously.
While locked the display shows "no ACCESS"
ON = treadmill is unlocked / accessible (standard)
OP41 Locking and unlocking the manual mode
OFF = manual mode is locked / not accessible
ON = manual mode is unlocked / accessible (standard)
OP42 Locking and unlocking the profile mode
OFF = profile mode is locked / not accessible
1 ... 6 = profile mode is unlocked / accessible up to the selected profile
number standard: 6
Example: Selected profile number = 3: The profiles 1-3 can be
selected, the profiles 4 – 6 cannot be selected
OP43 Locking and unlocking the cardio mode
OFF = cardio mode is locked / not accessible
ON = cardio mode is unlocked (standard)
OP44 Locking and unlocking the test mode
OFF = test mode is locked / not accessible
1 ... 94 = test mode is unlocked / accessible up to the selected test
profile number standard: 24
selected test number = 5: The test profiles 1-5 can be selected, the
test profiles 6 – 94 cannot be selected
OP45 Report mode display “Index“
0 = Display alternates (default)
1 = Display permanently in MET
2 = Display permanently in kJ
3 = Display permanently in Watt
After total switch off the default value 0 will be valid again.
OP46 Report mode display “Elevation“
in profile
mode and test mode
0 = Display alternates (default)
1 = Display permanently in % or degree (°), depending on OP14
2 = Display permanently in “Step“
After total switch off the default value 0 will be valid again.
OP47 Sustain values in display resp. automatic “Reset“ OFF = Display values are deleted after pressing START again or
automatically 2 minutes after having pressed STOP (default)
ON = Display values will be continued (added) after pressing START
again and will not be automatically deleted by pressing STOP.
Display values can only be deleted by pressing the STOP key twice
(time, distance, energy).