Quality Control
Controls (also known as “quality control materials”) are solutions for which an expected
analyte concentration range has been established. Controls are tested to check the
performance of your test system: CardioChek PA analyzer, MEMo Chip, and PTS Panels
test strips. Use quality control materials provided by PTS Diagnostics.
Refer to the range card provided with the controls or visit
http://www.ptsdiagnostics.com for control specifications.
Healthcare professionals should follow their facility’s guidelines and policies regarding
quality assurance and the use of quality control materials.
Quality control materials should be run:
•
With each new shipment
•
With each new lot number
•
According to state, local, and federal regulations
Important:
Check the expiration date printed on the control bottles. Do not use control
solutions that have expired.
For performing a quality control test, see the instructions below.
To perform a control test you need:
•
CardioChek PA analyzer
•
PTS Panels test strips
•
Quality control materials
•
Quality control instructions
•
Quality control range card
Quality Control Testing
5
How to Run a Quality Control Test
Refer to the instructions for use provided with your quality control materials.
If Quality Control Results Are Not in Range
IMPORTANT: Patient tests should not be performed until control results are
within range.
1. Ensure test strip slot area is clean.
2. Make sure neither the test strips nor the controls
are past the expiration date printed on the label.
3. Make sure the MEMo Chip matches the test strip lot.
4. Repeat the test again using fresh materials.
5. Call Customer Service for assistance.