Test results are automatically stored in the CardioChek PA analyzer’s memory. The
analyzer can store up to 30 results of each chemistry and 10 results for control tests. The
analyzer allows review of the results in order from the most recent to the oldest. Each
result is displayed with time and date. Results stored in memory are not deleted when
the batteries are changed.
How to Review Results Stored in Memory:
1. Press either button to turn on the analyzer. If the display reads INSTALL MEMO
CHIP, go to Step 2. If the display reads INSERT STRIP, press Enter.
2. Press Next until MEMORY is displayed.
3. Press Enter. CHEM is displayed.
4. Press Enter, then Next to select the desired chemistry.
Note:
Until the chemistry has been run at least once, the test name is
not displayed.
5. Press Enter to view the test result including time and date.
a.
To recall control results, press Next until EXIT is displayed. Press Enter. Press
Next until CONTROL is displayed.
b.
Press Enter when the desired control test is displayed.
c.
For example, to review lipid panel results, from the CHEM display, press
Next until LIPIDS is displayed, then Enter. The time and date will be
displayed. Press Enter when the desired test time and date is displayed.
Press Next to scroll through results.
6. To exit, press Next until the display reads EXIT, then press Enter. Repeat this step
until you return to RUN TEST.
Memory
7
Quick Reference - How to Run a Test
1. Press either button to turn the analyzer on.
2. Remove the MEMo Chip from the box of test strips.
3. Insert the MEMo Chip in the port at the top of the
analyzer with the lot code number facing down.
4. When INSERT STRIP is displayed, remove a test strip
from the vial and immediately replace the cap.
5. Insert the strip. Ensure that the test strip is inserted
fully and the display reads APPLY SAMPLE.
6. Obtain a blood drop following the correct
technique. (If venous blood is used, collect in an
EDTA or heparin tube. Invert gently 5-7 times to
mix completely. Immediately collect sample with
capillary tube or precision pipet and dispense
correct volume* as specified in test strip instructions
for use (package insert) on to the test strip.)
7. Hold the capillary tube by the bulb and position
above the blood application window on the test
strip. Use care to avoid touching the test strip with
the capillary tube. Squeeze the bulb gently to
deposit the entire sample on the strip.
8. Once the sample is applied, results will appear on
the analyzer display in about 90 seconds depending
on type of test strip.
9. Remove test strip and
dispose of properly.
10. If the analyzer is idle for more than 3 minutes, it will
count down 10 seconds and automatically turn off.
CHOL
mg/dL
215
* Refer to each specific test strip package insert for sample volume and sample application instructions.
Each test strip is for a single test only. Do not reuse the test strips. Use a new test strip each time you test.