DIADOS E
T11035
Operating Manual
– Putting the System into Operation
24
D616.131.00/10 en
When used with X-ray units for radiation qualities
according to DIN EN 61267 (RQR 2 to RQR 10 and
RQA 2 to RQA 10) in range 40 to 150 kV:
dose or dose rate during radiography and
fluoroscopy
exposure time or irradiation time
dose per pulse and number of pulses
When used with mammography units for radiation
qualities according to DIN EN 61267 (RQR-M 1 to
RQR-M 4 and RQA-M 1 to RQA-M 4) as well as
Mo-Anode 25 µm Rh (+ 2 mm Al),
W-Anode 50 µm Rh (+ 2 mm Al) and
Rh-Anode 25 µm Rh (+ 2 mm Al)
W-Anode 50 µm Ag (+ 2 mm Al)
W-Anode 0.7 mm Al (+ 2 mm Al),
in range 25 to 45 kV:
dose or dose rate
exposure time or irradiation time
dose per pulse and number of pulses
When used at dental X-ray units in the range of
40 kV to 90 kV with filter 8.5 mm Al or 8.5 mm Al +
0.5 mm Cu:
dose or dose rate
exposure time or irradiation time
dose per pulse and number of pulses
When used with CT systems for radiation qualities
according to DIN EN 61267 (RQT 8 to RQT 10) in
range 100 to 150) kV:
dose length product or dose length product
rate
exposure time or irradiation time
dose per pulse and number of pulses