4.3.4 Display during the Measurement
Normal mode:
Depending on your selection, the current, the
charge or the measuring time will be displayed with
the
ALT
key. With
ALT
you can toggle between the
displayed parameters during and after the meas-
urement.
Pulsed mode:
Depending on your selection, the charge, the
charge per pulse or the number of pulses will be
displayed with the
ALT
key. With
ALT
you can tog-
gle between the displayed parameters during and
after the measurement.
Digital resolution:
If the resolution is insufficient, one or two asterisks
(* or **) are displayed to the left of the measured
value. If the digital resolution of the display is below
1%, two asterisks ** will appear, with a resolution
between 0.5% and 1% one asterisk * appears; if the
digital resolution is greater than 0.5% no asterisks
are displayed.
4.3.5 Ending the Measurement
A measurement initiated with Autostart ends auto-
matically when no measuring signal is detected for
more than the measuring delay time.
A manually initiated measurement is stopped with
START/STOP.
4.3.6 Display at End of Measurement
Normal mode:
After the measurement, you can toggle between the
measured values with the
ALT
key. The current is
given as 0.00 A.
Pulsed mode:
After the measurement, you can toggle between the
measured values with the
ALT
key.