DIADOS E
T11035
Operating Manual
– Description of the Device
12
D616.131.00/10 en
1.2 Connection Panel
DIADOS E has connectors for the power supply
(
9 VDC
) and for the detector (
DETECTOR
) and an
RS232 interface (
SERIAL
) (see Figure 2).
As a general rule, the connector for the detector
should be sealed with the enclosed cap when not in
use.
Figure 2:
DIADOS E connection Panel
1.3 Display
While a measurement is in progress, the current
dose/dose rate is indicated in intervals of 0.5 s.
The
ALT
key can be pressed to toggle between the
different readouts during a measurement.
Figure 3:
DIADOS E display
1
Resolution:
" " greater than 0.5 %,
"*" greater than 1 %,
"**" less than 1 %
2
Unit C, Gy or R; displayed in reverse video in
pulsed mode
3
Application:
RAD
– convential radiology,
DENT
– dental application,
CT
– CT application or
MamX/X
– mammography
(MamM/M, MamM/R, MamW/A,
MamW/R, MamR/R, MamWAg)
4
Measuring range: Low, Med, High or Auto
5
Measuring value
6
Measurement status:
RDY
– ready to take a measurement,
RUN
– measurement in progress,
STA
– measurement started with START/STOP,
STP
– measurement stopped with
START/STOP,
TST
– measurement stopped by timer
7
Filter:
"Unatt."
(without
additional
filtration)
or
"Att. " (with additional filtration)
8
Battery status: full/depleted
DETECTOR
SERIAL
1 2 3 4
5 6 7 8