DIADOS E
T11035
Operating Manual
– Putting the System into Operation
D616.131.00/10 en
21
2.3.2 Use with CT units
For this application a CT chamber (TM30009,
TM30017 or TM77336) with CT adapter (T16018)
has to be used.
Connecting the CT chamber
The CT adapter is connected to the
DETECTOR
socket at the connection panel of the DIADOS E
(refer to Figure 2); the CT chamber is connected to
the CT adapter. For connecting the chamber, the
DIADOS E may be switched on or off.
NOTE
An extension cable used shall be connected
between DIADOS E and the CT adapter (con-
nection system W). The overall length of the
detector cable with extension cable shall not
exceed 20 meters.
Also refer to the user manual supplied with the
CT chamber.
NOTE
When used with CT systems:
In conjunction
with the multi-range power supply, power surges
may inadvertently trigger the device (starting a
measurement). This phenomenon does not occur
when the device is exclusively powered from
batteries. In CT applications (CT chamber and
CT junction box) operate DIADOS E only on
battery power to avoid inadvertent measure-
ments.
Checking the high voltage of the CT adapter
Push the TEST key at the adapter and hold it de-
pressed. The LED should illuminate and go out
within 2 s. If the LED does not light up or go out, the
high voltage supply is defective and correct meas-
urements cannot be performed. In this case the CT
adapter should be returned to the manufacturer for
repair.
WARNING
Electricity is a source of risk, particularly when
the product is not in perfect operating condition
or when it is operated inappropriately.
Patient Hazard! Shock Hazard!
Only qualified staff is authorized to open the CT
adapter (electric shock hazard!). Never attempt
to repair a defective CT adapter.
2.4 Positioning DIADOS E
Position DIADOS E such that the control panel is
within easy reach and that the display can be read
off.