8.5 Maintenance of Officially
Calibrated Dosemeters
This paragraph concerns DIADOS E users in
countries where an official calibration is re-
quired, e.g. Germany.
In case of officially calibrated dosemeters, the user
is not allowed to open the instrument and to per-
form manipulations that may change the measuring
value. Therefore the housing is sealed with official
stickers. If one or more of these stickers are violat-
ed, then the official calibration is not longer valid.
If a dosemeter is not working correctly, then it may
be sent for repair to a service station that is ap-
proved by the authorities. After repair the doseme-
ter must be officially calibrated again. However, it
may be used until this new official calibration can be
done.
If the repair has been performed by a non-approved
service station, then the dosemeter may only be
used for applications that are under the official cali-
bration procedure if it has been officially re-
calibrated.
8.6 Calibration
When using not officially calibrated dosemeters, we
recommend having the system calibrated every 2
years at PTW-Freiburg or by specially trained staff
on site.
8.7 Rechargeable Batteries
DANGER
Improper handling of batteries.
Explosion Hazard!
Do
not
throw the rechargeable batteries into a
fire.
WARNING
Improper handling of batteries.
Bodily injury!
If electrolyte leaks from the battery, avoid contact
with skin or clothes. Immediately clean skin or
clothes that have come in contact with electrolyte
with water. In case of eye contact, immediately
rinse eyes with water and consult doctor.
Do not open rechargeable batteries.
WARNING
Improper handling of batteries.
Risk of Fire or Skin Burns!
Never short the rechargeable batteries together.
WARNING
Improper handling of batteries.
Bodily injury! Equipment damage!
Equipment damage - Do not open rechargeable
batteries.
Never
expose rechargeable batteries
to water. Store and use the rechargeable batter-
ies at the ambient conditions (temperature, rela-
tive humidity) specified for the device.