MULTIDOS

RS232 Interface

Application Dual Channel Dosemeter

10

D177.131.2/0

7.2 Data telegrams

7.2.1  Read measured values

The response to a data telegram consists of various fields, separated by semicolons. The fields contain pa-
rameters and text.
The response to the D telegram is:
D

m

;

ttttt.n

s;

sss

;

FL

;

O

;

L

;

M

;

±m.mmmE±ee

;

a

;

±m.mmmE±ee

;

a

;

·####.#

;

bbbbb

The measured values and resolution indicators are that of the actual measuring mode and are transmitted
channel 1 first, followed by channel 2. The ratio is always transmitted, even if absolute values are displayed
(see below).

7.2.2  Description of parameters

m

Actual measuring mode. Valid values: 0 

Þ

 Dose or Charge, 1 

Þ

 Doserate or Current

ttttt.n

Elapsed measurement time. Consists of seven characters. Small numbers are right justified,
with leading spaces. Parameter n can have the values of 0 or 5.
If the maximum measurement time of 64800 seconds is exceeded,  OL•••••  will be transmit-
ted.

sss

Measurement status. Valid values: RES, STA, HLD, INT, RUN, NUL, ERR.

FL

Global flags. A two digit number, whose set bits, when converted to binary representation,
signal the following error:
.0 

Þ

 any of both channels is overloaded at the moment

.1 

Þ

 math error in any channel

.2 

Þ

 MULTIDOS data acquisition error

.3 

Þ

 MULTIDOS HVerror at the moment

.4 

Þ

 overload in any of both channels since the start of a dose measurement

.5 

Þ

 MULTIDOS HV error since the start of a dose measurement

.0 is the least significant bit

O

Overload flags of the doserate / current measurement. One digit, whose set bits, when con-
verted to binary representation, signal the following errors:
.0 

Þ

 overload in channel 1

.1 

Þ

 overload in channel 2

L

Latched overload flags of the dose / charge measurement. One digit, whose set bits, when
converted to binary representation, signal the following errors:
.0 

Þ

 overload in channel 1 since the start of a dose measurement

.1 

Þ

 overload in channel 2 since the start of a dose measurement

M

Math error flags of the actual measurement mode. One digit, whose set bits, when converted
to binary representation, signal the following errors:
.0 

Þ

 math error in channel 1

.1 

Þ

 math error in channel 2

±m.mmm

Mantissa of the measured value. This field consists of six characters. Smaller numbers are
right justified, with leading spaces. Positive numbers use a space character instead of the
minus sign.
If the limits of representation (±999.9E+20) are exceeded, +0L•••  or  -0L•••  will be transmit-
ted, depending on the sign.

E±ee

Exponent of the measured value. Two digit number, with leading zeroes if necessary. The sign
is always transmitted, either as + or -.
If the limits of representation are exceeded, •••• will be transmitted.

a

Resolution of the measured value. One digit. Valid values:

Þ

 resolution below 1% (conforming to ** on the measuring screen)

Þ

 resolution below 0.5% (conforming to *)

Þ

 resolution is 0.5% or better

Summary of Contents for MULTIDOS

Page 1: ...User Manual MULTIDOS RS232 Interface D177 131 2 0 2003 07 Di Se...

Page 2: ...MULTIDOS RS232 Interface 2 D177 131 2 0 For your notes...

Page 3: ...sic Settings 9 7 2 Data telegrams 10 7 3 Correction Telegrams 11 7 4 Calibration Factors 11 8 APPLICATION MULTI CHANNEL DOSEMETER 12 8 1 Basic Settings 12 8 2 Data telegrams 12 8 3 Correction Telegram...

Page 4: ...th user manuals are applicable and must be ob served In addition to these safety statements please note that devices connected to the interface described in this manual must fulfill the requirements o...

Page 5: ...RTS out Pins 4 and 6 are internally connected By this an external DTR signal may be returned as DSR to the external device Either RTS CTS hardware handshake or no hand shake may be used 4 Principles o...

Page 6: ...hannel L Linear Array Wait 5 seconds after sending Aa Aa Aquit Restart the current application Wait 5 seconds after sending Aquit Aquit BRbbbbb Set Read baud rate Parameter bbbbb 04800 09600 19200 384...

Page 7: ...Set1 of application dual channel is write protected 4 radiological unit is Roentgen 5 reference temperature for air density correction is 22 C 6 HV error 7 ME 48 is connected to MULTIDOS 0 is the leas...

Page 8: ...ore characters Repeat this telegram three times maximum if you do not receive the expected response or no response at all 2 For all telegrams if the expected response is not received MULTIDOS is set u...

Page 9: ...ded by channel 1 Qx Rxy Set Read range settings Parameter x setting of channel 1 Parameter y setting of channel 2 Valid values of x and y L Low H High keep current setting Valid values of the response...

Page 10: ...ince the start of a dose measurement 5 MULTIDOS HV error since the start of a dose measurement 0 is the least significant bit O Overload flags of the doserate current measurement One digit whose set b...

Page 11: ...value is the resolution of the actual measuring mode for the active unit radiological or electrical and actual measuring range of that channel The number of digits following the decimal point may vary...

Page 12: ...s FL global measurement flags as Overload in any channel HVerror etc Two digit number lm number of maximum channel of group 1 6 marked on display by Two digit number um number of maximum channel of gr...

Page 13: ...overall correction factor k k 0 500 2 000 KKn nnn 8 4 Calibration Factors Telegram Description Response CRsFccf Set Read the calibration factor f of channel cc of set s CRsFccf CRsIx Set Read the Is c...

Page 14: ...he XMODE is activated with XM1 If it is not ac tivated you get E01 back as the telegram response It is only possible to select a set if the set is calibrated or if it is possible to calibrate the MULT...

Page 15: ...ted m mmm mmm Mantissa of the measured value This field consists of 9 characters Smaller numbers are right justified with leading spaces Positive numbers use a space character instead of the minus sig...

Page 16: ...cc f Set Read the radiological calibration factor for channel cc in set s XR s cc m m mmmE ee XRF s Set Read the flag for set s for radiologically calibrated uncalibrated 0 not calibrated 1 calibrate...

Page 17: ...r the reference channel M for the monitor signal V1 addresses the channel which measures the 900 volts V4 that to measure the 400 volts m Measuring mode Valid values 0 Dose or Charge 1 Doserate or Cur...

Page 18: ...f bbbbb 10 2 2 Single channel data Telegram cc Response Notes Dcc 01 47 Without reference Dcc m ttttts sss m mmmE ee f FL bbbbb 01 47 With reference Dcc m ttttts sss m mmm f FL bbbbb 1 R or M Dcc m tt...

Page 19: ...thout reference 26 0 47 13 5 characters 642 characters 10 2 4 Measurement resolution Telegram Response DRcc DRcc0 mmmE ee Reads the absolute resolution of channel cc Notes 1 This is the resolution of...

Page 20: ...is active 0 is the least significant bit aa can take values from 00 to 63 SETAaa 11 2 Control Telegrams Telegram Description Response NEW Reinitialize measurement NEW NULE Read the channels only acti...

Page 21: ...5 channel 6 0 is the least significant bit OO can take values from 00 to 63 LL Latched overload flags of the dose or charge measurement 2 digit whose set bits when con verted to binary representation...

Page 22: ...am Description Response CRsAaa Set Read the status of the channels of set s aa is a 2 digit decimal number whose set bits when converted to binary representation signal the following status 0 channel...

Reviews: