background image

MULTIDOS

RS232 Interface

General Telegrams

6

D177.131.2/0

5 General Telegrams

These telegrams are valid for all MULTIDOS applications.

5.1 Device Identification

Telegram

Description

Response

PTW

Read device information. x.xx is the firmware version, u describes the ra-
diological unit

MULTIDOS•x.
xxu

SER

Read MULTIDOS serial number

SERxxxxxx

5.2 Keyboard

Telegram

Description

Response

K

x

Disable or reenable the MULTIDOS keyboard. Parameter 

x

: 0 = disabled, 1

= enabled

Kx

5.3  Basic Device Settings

Telegram

Description

Response

L

x

Set/Read language setting. Parameter 

x

: E = Englisch, D = German

Lx

I

dddd

Set/Read interval time. Parameter 

dddd

: 0006 up to 9999

Idddd

A

a

Set/Read the MULTIDOS application. Parameter 

a

: A = Afterloading, C =

Constancy Check, D = Dual Dosemeter, M = Multi Channel, L = Linear
Array.
Wait 5 seconds after sending A

a

Aa

Aquit

Restart the current application. Wait 5 seconds after sending Aquit

Aquit

BR

bbbbb

Set/Read baud rate. Parameter 

bbbbb

: 04800, 09600, 19200, 38400

BRbbbbb

5.4  Controlling a Measurement

Telegram

Description

Response

STA

Equivalent to <STA> key

STA

INT

Equivalent to <INT> key

INT

RES

Equivalent to <RES> key

RES

HLD

Equivalent to <HLD> key

HLD

NUL

Equivalent to <NUL> key. Response NUL or E06 is send after 28 seconds

NUL

M

x

Set/Read measuring mode. Parameter 

x

: 0 = Dose, 1 = Doserate

Mx

S

Read measurement status.
Valid values of parameter 

sss

: RES, STA, HLD, INT, RUN, NUL, ERR

Ssss

5.5 Error Codes

Error code

Description

E01

Unknown command or illegal parameter

E02

Command in wrong context (STA for doserate measurement, etc.)

E03

Command not allowed at the moment (MULTIDOS is in a menu)

E06

Error during zeroing

E07

Answer could not be sent (transmission buffer full)

E09

Error during write to EEPROM

E10

Parameter out of limits

Summary of Contents for MULTIDOS

Page 1: ...User Manual MULTIDOS RS232 Interface D177 131 2 0 2003 07 Di Se...

Page 2: ...MULTIDOS RS232 Interface 2 D177 131 2 0 For your notes...

Page 3: ...sic Settings 9 7 2 Data telegrams 10 7 3 Correction Telegrams 11 7 4 Calibration Factors 11 8 APPLICATION MULTI CHANNEL DOSEMETER 12 8 1 Basic Settings 12 8 2 Data telegrams 12 8 3 Correction Telegram...

Page 4: ...th user manuals are applicable and must be ob served In addition to these safety statements please note that devices connected to the interface described in this manual must fulfill the requirements o...

Page 5: ...RTS out Pins 4 and 6 are internally connected By this an external DTR signal may be returned as DSR to the external device Either RTS CTS hardware handshake or no hand shake may be used 4 Principles o...

Page 6: ...hannel L Linear Array Wait 5 seconds after sending Aa Aa Aquit Restart the current application Wait 5 seconds after sending Aquit Aquit BRbbbbb Set Read baud rate Parameter bbbbb 04800 09600 19200 384...

Page 7: ...Set1 of application dual channel is write protected 4 radiological unit is Roentgen 5 reference temperature for air density correction is 22 C 6 HV error 7 ME 48 is connected to MULTIDOS 0 is the leas...

Page 8: ...ore characters Repeat this telegram three times maximum if you do not receive the expected response or no response at all 2 For all telegrams if the expected response is not received MULTIDOS is set u...

Page 9: ...ded by channel 1 Qx Rxy Set Read range settings Parameter x setting of channel 1 Parameter y setting of channel 2 Valid values of x and y L Low H High keep current setting Valid values of the response...

Page 10: ...ince the start of a dose measurement 5 MULTIDOS HV error since the start of a dose measurement 0 is the least significant bit O Overload flags of the doserate current measurement One digit whose set b...

Page 11: ...value is the resolution of the actual measuring mode for the active unit radiological or electrical and actual measuring range of that channel The number of digits following the decimal point may vary...

Page 12: ...s FL global measurement flags as Overload in any channel HVerror etc Two digit number lm number of maximum channel of group 1 6 marked on display by Two digit number um number of maximum channel of gr...

Page 13: ...overall correction factor k k 0 500 2 000 KKn nnn 8 4 Calibration Factors Telegram Description Response CRsFccf Set Read the calibration factor f of channel cc of set s CRsFccf CRsIx Set Read the Is c...

Page 14: ...he XMODE is activated with XM1 If it is not ac tivated you get E01 back as the telegram response It is only possible to select a set if the set is calibrated or if it is possible to calibrate the MULT...

Page 15: ...ted m mmm mmm Mantissa of the measured value This field consists of 9 characters Smaller numbers are right justified with leading spaces Positive numbers use a space character instead of the minus sig...

Page 16: ...cc f Set Read the radiological calibration factor for channel cc in set s XR s cc m m mmmE ee XRF s Set Read the flag for set s for radiologically calibrated uncalibrated 0 not calibrated 1 calibrate...

Page 17: ...r the reference channel M for the monitor signal V1 addresses the channel which measures the 900 volts V4 that to measure the 400 volts m Measuring mode Valid values 0 Dose or Charge 1 Doserate or Cur...

Page 18: ...f bbbbb 10 2 2 Single channel data Telegram cc Response Notes Dcc 01 47 Without reference Dcc m ttttts sss m mmmE ee f FL bbbbb 01 47 With reference Dcc m ttttts sss m mmm f FL bbbbb 1 R or M Dcc m tt...

Page 19: ...thout reference 26 0 47 13 5 characters 642 characters 10 2 4 Measurement resolution Telegram Response DRcc DRcc0 mmmE ee Reads the absolute resolution of channel cc Notes 1 This is the resolution of...

Page 20: ...is active 0 is the least significant bit aa can take values from 00 to 63 SETAaa 11 2 Control Telegrams Telegram Description Response NEW Reinitialize measurement NEW NULE Read the channels only acti...

Page 21: ...5 channel 6 0 is the least significant bit OO can take values from 00 to 63 LL Latched overload flags of the dose or charge measurement 2 digit whose set bits when con verted to binary representation...

Page 22: ...am Description Response CRsAaa Set Read the status of the channels of set s aa is a 2 digit decimal number whose set bits when converted to binary representation signal the following status 0 channel...

Reviews: