NOMEX
Operating Manual - Putting the NOMEX Multimeter into Service
28
D909.131.00/02 en
Measurement on Dental Panoramic Units in
the Presence of Rotating Radiation:
With this type of equipment it must also be
assured that the NOMEX multimeter is always
fully irradiated statically when the rotation cannot
be avoided, i.e., the NOMEX multimeter must fol-
low the rotating movement. For this purpose, you
can use the NOMEX multimeter holder T20016,
for example (refer to section