D909.131.00/02 en
5.4.2 Determining the Dose Unit
The dose or dose rate can be displayed in Gray
[Gy] or Roentgen [R] units.
•
To display the units [Gy] or [Gy/s], select menu
View
→
Dose unit
→
[Gy]
.
•
To display the units [R] or [R/s], select menu
View
→
Dose unit
→
[R]
.
NOTE
The units [R] and [R/s] are not available on
legally calibrated devices.
5.4.3 Determining the Type of Tube
Voltage Displayed
After the measurement, the following values are
determined for the tube voltage:
– mean peak voltage (mean kVp)
The mean kVp corresponds to the mean value
of the upper 10 % of the highest measured
tube voltages.
– maximum peak voltage (maximum kVp)
The maximum kVp corresponds to the highest
measured tube voltage that has been recorded
5 times.
– practical peak voltage (PPV)
The PPV is calculated according to the algo-
rithm of IEC 61676.
One of these values can be selected for display:
•
To display the mean peak voltage, select menu
View
→
Tube Voltage
→
Mean kVp
.
•
To display the maximum peak voltage, select
menu
View
→
Tube Voltage
→
Max. kVp
.
•
To display the practical peak voltage, select
menu
View
→
Tube Voltage
→
PPV
.
or
•
Continue clicking the appropriate row in the
measuring window until the desired value is